user manual

6.4.5
6 - 93
(f) Drag the cursor with the mouse and align it with the part to be measured, then click
Enter button. Angle measurement can thus be done as in Fig. 6.4-22. Calibration is
not performed in the case of angle measurement.
Fig. 6.4-22 Angle Measurement Result
(2) Successive Manual Measurement
(a) By selecting Multiple under manual measurement mode and under the same
measurement parameters, successive manual measurement can be carried out.
Click Set button on the CD Measurement window and the measurement cursors
appear. In successive measurement, X and Y directions cannot be measured
simultaneously, so you must specify a measuring direction.
(b) Drag the measurement cursor box with the mouse as in the manual measurement,
and align the cursor with the part to be measured. Click Enter button and the
measured result appears.
(c) Click Set button in this status to display the measurement cursors, then align the
cursor with the next part to be measured. Click Enter button and the 2nd measured
result will appear after the 1st measured result.
(d) In the same manner, click Set button, align the measurement cursors, and click Enter
button to achieve successive measurement. In manual mode, a maximum of 8
successive measurements can be achieved.
(e) Successive measurement is also possible in the angle measurement mode.
Successive measurement by changing X, Y or Angle midway as in Fig. 6.4-23 is also
feasible. In the case of Angle, numerics will be displayed in the part where the
circular cursor appears.