User Manual

HOBO® 4-Channel Thermocouple
Data Logger (UX120-014M) Manual
17031-D
The HOBO 4-Channel Thermocouple data logger records temperature in indoor environments
using up to four J, K, T, E, R, S, B, or N type thermocouple sensors (sold separately). The logger
also has an internal 10K thermistor for cold-junction compensation of the thermocouple
output. The four thermocouple channels have a 20-bit resolution while the internal 10K
thermistor has a 12-bit resolution. Using HOBOware®, you can easily configure the logger alarm
to trip for specific high or low temperatures. Or, you can set up burst logging in which the
logger records data at a different interval during certain conditions. The logger can also
calculate minimum, maximum, average, and standard deviation statistics. This easy-to-use data
logger has a built-in LCD screen to monitor temperature, logging status, battery use, and
memory consumption in between readouts. An optional protective case (CASE-4X-2) is also
available for this logger.
Specifications
Thermocouple
Type Range Accuracy Resolution
J -210° to 760°C
(-346° to 1,400°F)
±0.6°C (±1.08°F) ± thermocouple
probe accuracy
0.03°C (0.06°F)
K -260° to 1,370°C
(-436° to 2,498°F)
±0.7°C (±1.26°F) ± thermocouple
probe accuracy
0.04°C (0.07°F)
T -260° to 400°C
(-436° to 752°F)
±0.6°C (±1.08°F) ± thermocouple
probe accuracy
0.02°C (0.03°F)
E -260° to 950°C
(-436° to 1,742°F)
±0.6°C (±1.08°F) ± thermocouple
probe accuracy
0.03°C at (0.05°F)
R -50° to 1,550°C
(-58° to 2,822°F)
±2.2°C (±3.96°F) ± thermocouple
probe accuracy
0.08°C (0.15°F)
S -50° to 1,720°C
(-58° to 3,128°F)
±2.2°C (±3.96°F) ± thermocouple
probe accuracy
0.08°C (0.15°F)
B 550° to 1,820°C
(1,022° to 3,308°F)
±2.5°C (±4.5°F) ± thermocouple
probe accuracy
0.1°C (0.18F)
N -260° to 1,300°C
(-436° to 2,372°F)
±1.0°C (±1.8°F) ± thermocouple
probe accuracy
0.06°C (0.11°F)
Internal 10K Thermistor (Temperature)
Range -20° to 70°C (-4° to 158°F)
Accuracy ±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A
Resolution 0.024°C at 25°C (0.04°F at 77°F); see Plot A
Drift <0.1°C (0.18°F) per year
Logger
Logger Operating Range Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing);
Launch/Readout: 0° to 50°C (32° to 122°F) per USB specification
Logging Rate 1 second to 18 hours, 12 minutes, 15 seconds
Logging Modes Fixed interval (normal), burst, or statistics
Memory Modes Wrap when full or stop when full
Start Modes Immediate, push button, date & time, or next interval
Stop Modes When memory full, push button, or date & time
Restart Mode Push button
Time Accuracy ±1 minute per month at 25°C (77°F), see Plot B
Power Source Two AAA 1.5 V alkaline batteries, user replaceable, and USB cable
Battery Life 1 year, typical with logging rate of 1 minute and sampling interval of
15 seconds or greater
Memory 4 MB (1.6 million measurements, maximum)
HOBO 4-Channel
Thermocouple
Data Logger
UX120-014M
Included Items:
Command™ strip
Double-sided tape
Hook & loop strap
Two AAA 1.5 V alkaline
batteries
Required Items:
HOBOware 3.5 or later
USB cable (included with
software)
Thermocouple sensor type
J, K, T, E, R, S, B, or N
Accessory:
Protective case (CASE-4X-2)

Summary of content (8 pages)