User's Manual

Specifications
Overall system dimension
Size
76"X24"X35"
Weight - Full loaded condition
1,000lbs (454kg)
Weight - Empty rack
≤730lbs (331kg)
Analyzer
Size
22" x 13" x 6"
Weight
≤ 27lbs(12kg)
Power Requirements
Operating Voltage
DC 24V / 2.5A
Power Consumption
Normal condition
1125 Watt
Maximum condition
<1400 Watt
Outputs
Visual
15.6" wide screen monitor with capacitive touch interface
Display gas reading, alarm & instrument status realtime base
Provied realtime trend chart and gas event snapshoot trend chart.
Event logging including maintenance/instrument fault and gas alarm
Muiltipe searching option for the event
LED indicator: Normal (Green), Alarm 1 (Red), Alarm 2 (Red), Fault (Yellow)
Digital Communications
TCP/IP and RTU Modbus, OPC (Option)
Security
Role-based access control
Support HTTPS
Certification and Specification
EN 50270 and EN61010-1
UL/IEC 61010-1
RED, FCC for RFID
Performance
Refer to gas list table
Transport System
Flow Rate
2.4 LPM
Transport Time
Less than 45 seconds upto 328ft with thin wall tube (0.190"ID)
Less than 40 seconds upto 400ft with medium wall (0.156"ID)
Sample Line Tubing
1/4 in. (6.35 mm)O.D. x 0.190 in. (4.83 mm) I.D. (Thin wall)FEP Teflon
or 1/4 in. (6.35 mm) O.D. x 0.156 in. (3.9 mm) (Medium wall)
Tubing Length
Up to 328ft (100m) with thin wall; Up to 400ft (122m) with medium wall
Exhaust Line Tubing
1/2 in. (12.7 mm) O.D. x 3/8 in.
(9.5 mm) I.D. Teflon tubing,
Exhaust Length
Up to 50 ft. (15 m) maximum
Operating conditions
Temperature
59°F to 95°F (15°C to 35°C)
Humdity
20-80% RH
Alititude
1000 ft. (305 m) to 6000 ft. (1829 m) above sea level
Pollution Degree
2
Wiring Requirement
Power
Singe phase power, Minimum 14 AWG
Digital
CAT5 Cable or equivalent: RJ45 connector
© 2020