User Manual

Commissioning
SmartRadar FlexLine Part No.: 4417.762_Rev07
6 - 76 Service Manual
reflections from the verification pins are stronger and recognized by the
SmartRadar FlexLine.
FIGURE 6-15 Verification pin measuring position
To measure the position of the verification pins, the SmartRadar
FlexLine must be turned 90 degrees as follows:
Release the coupling nut of the SmartRadar FlexLine housing.
Lift the SmartRadar FlexLine housing from tank separator.
Rotate the SmartRadar FlexLine for 90°. This can only be done in
one direction because of the locking pin.
Place SmartRadar FlexLine on tank separator; mind locking pin.
Secure the coupling nut of the SmartRadar FlexLine housing.
Then issue the command:
[Measure verification pins] (on SmartView:
Calibrate v-pins).
When the verification pin measurement is completed, the SmartRadar
FlexLine housing must be turned back 90° (follow above procedure in
reverse order).
The verification-pin compensation must be enabled by the appropriate
switch. In Engauge, that is (depending on installed TII-XR firmware and
board descriptor):
the second position of correction field:
(up to TII-XR firmware version A1130 and board descriptor V4)