User Manual

Commissioning
SmartRadar FlexLine Part No.: 4417.762_Rev07
6 - 82 Service Manual
6.6.3 Relay Configuration
6.6.3.1 Jumper Settings
At installation, each individual relay contact was configured as required
with the hardware jumpers JPx0, where x = Relay 1 to 4 respectively.
See FIGURE 6-20.
ESF07-0002
FIGURE 6-20 The relays’ hardware jumpers
NOTE: In the Commissioning stage, no jumper setting can
be changed without breaking the compartment screw
sealing.
6.6.3.2 Relay Mode
Each individual relay can be set to be energized or de-energized during
operation, by setting the
[Relay Mode] entity to <Energized> or <De-energized>
respectively.
If the
[Relay Mode] entity is set to <Energized>, the relay coil will be
energized when the relay state is
<Deactivated>, and the relay coil will be
de-energized when the relay state is
<Activated>.
If the
[Relay Mode] entity is set to <De-Energized>, the relay coil will be de-
energized when the relay state is
<Deactivated>, and the relay coil will be
energized when the relay state is
<Activated>.
Ryx_b
Ryx_a
JPx0
n.o. n.c.