Commissioning
SmartRadar FlexLine Part No.: 4417.762_Rev07
6 - 160 Service Manual
6.11.4.4.2 Engauge * Vapour temperature * Tab
NOTE: Some settings for vapour temperature calculations
are shared via the product temperature settings.
The following entities can be set by Engauge or SmartView for a
specific functioning of the FII-RTD module in an instrument. The
default values are available after initialization of the non volatile
memory; so check each entity for its correctness.
[MRT length table]
[Only for MRT]
Specifies whether a fixed
range of MRT resistors is
used (=
<F> ) or user
configured lenghts (=
<T>).
Fixed lengths are:
0.65, 1.25, 1.95, 2.85, 4.15,
5.65
1 character
<F>
<T>
<F>
[MRT element length
used if MRT length table is 'T']
[Only for MRT]
The lengths of the MRTs
including anchor eye
floating point numbers:
<-x.x .. +x.x>
<0.0>
Name Explanation Value Range Default
[Gas immersion depth] The minimum required
distance below an element
before it is taken into
account in the average
vapour temperature
calculation
floating point number:
<-x.x .. +x.x>
<0.5>
[High High alarm]
[High Alarm]
[Low Alarm]
[Low Low alarm]
4 Thresholds for activating
a related alarm status in the
Primary Value
floating point numbers:
<-x.x .. +x.x>
<+1.0E22>
[Alarm test enable]
Enables (if activated, see
next listed entity) the
simulation of one of the 4
alarms for a minute, by
simulating the actual
measured Primary Value is
below or above the alarm
threshold.
<enable>
<disable>
<disable>
Name Explanation Value Range Default