User Manual
Commissioning
Part No.: 4417.762_Rev07 SmartRadar FlexLine
Service Manual 6 - 3
The HCI-BPM module supports 2 protocols:
The Enraf GPU protocol with its records and items (limited);
The Enraf FlexConn protocol with its so-called entities.
The module can communicate with:
880 CIU prime
858 CIU
780 SmartLink
6.2.2 Commissioning the HCI-BPM
For a correct functioning of the HCI-BPM module within an instrument
(gauge), the following entities can be set by using either Engauge or
SmartView.
☛ By using the following table, check each entity for correctness.
Name Value Range Default Value Explanation
[Baudrate] <1200>
<2400>
<4800>
<1200>
Communication speed
[BPM sensitivity] <1..8> <8> The sensitivity of the
receiver circuit
1 = weakest
8 = strongest
[Identification] 8 characters e.g.
<TANK1234>
<-------->
Name of a tank or
instrument
[GPU instrument address] <0..99> <0> The address of this
instrument for GPU
messages.
Note: Each instrument must
have a unique GPU
address.
[FlexConn instrument address] <0..1899> <0> The address of this
instrument for FlexConn
messages.
Note: Each instrument must
have a unique FlexConn
address.
[Level units] <meters>
<inches>
<feet>
<fractions>
<meters> The unit in which level-
related GPU records and
items are shown
[Temperature units] <celsius>
<fahrenheit>
<celsius> The unit in which
temperature-related GPU
records and items are
shown