6.4.5.2 Transducer Blocks
6.4.5.2.1 Introduction
Besides the Radar Level transducer block, up to 16 additional
transducer blocks can be added to the device via Wireless Builder. This
can easily be done by “drag and drop”.
The specific parameters of all available transducer blocks can be set via
Wireless Builder. This section describes how to use the Configuration
Form to configure the transducer blocks (channels).
Please refer to the Configuring Channels chapter in the OneWireless
manual. This Transducer Blocks section only explains the
additional or non-standard items.
6.4.5.2.2 Adding Transducer Blocks
In order to configure each channel, a user must add transducer blocks
to this device.
Expand the device template in the Library tree (click +), to see all
available transducer blocks.
Drag and drop the transducer block(s) that must be instantiated
(= made concrete) from the Library tree to the device in the Offline
tree.
NOTE: Each transducer block can only be added to the
instrument once.
After adding the relevant transducer block(s), load the device to
activate the transducer blocks.
NOTE: Please refer to Block instantiation in the OneWireless
Wireless builder User’s Guide for more information.
The following additional transducer blocks are available:
Transducer Block See section
Ambient temp -
Average product temperature -
Configurable Transducer block 1 6.4.5.2.5
Configurable Transducer block 2
Configurable Transducer block 3
Configurable Transducer block 4
Observed Density -
Product pressure -
Relay output 1 6.4.5.2.7
Relay output 2
Relay output 3
Relay output 4
Temperature Details 6.4.5.2.6
Wireless