Honeywell BW Solo User Manual
12
3.2. Status Indicator Icons
On-screen icons are used to indicate battery, bump, calibration and sensor status, as well as peak
reading. The table below details how and when each icon is displayed. The triangular red LED
located to the right of the screen illuminates to accompany these (it is active when the non-
compliance feature is on).
Icon
Explanation
The battery icon is not shown on the display unless the remaining power is one
day or less.
When the battery’s remaining life falls below 24 hours, the battery icon is
shown and there is a flash and beep every 60 minutes.
When the battery’s remaining life falls below 8 hours, the icon blinks on and
off every second, and there are two beeps and two flashes every 30
seconds.
When less than 12 hours remains before a bump test should be performed, this
icon is shown and the buzzer sounds once every hour. When the user has also
set the non-compliance option for the bump test reminder, the instrument flashes
once and beeps once every 10/30/60 seconds (the interval is set by the user, and
this beep function can be enabled/disabled).
The icon blinks once per second if there is a failed or skipped bump test, and the
instrument beeps three times every 60 minutes.
A bump test is required (and indicated by this icon) if:
The defined period between bump tests has been exceeded (bump test
overdue).
The sensor has failed a previous bump test.
Failed bump test notification is shown on the screen if a bump test is failed. In the
information menu, it shows that the bump test must be performed “NOW.”
Note: the sensor should be challenged on a regular basis (before each day’s use)
to ensure proper sensor performance.
This icon is shown when there is less than 12 hours remaining before a calibration
should be performed. The instrument beeps once every 60 minutes.
The icon blinks once per second if there is a failed or skipped calibration, and the
instrument beeps three times every hour if the non-compliance option is off. When
the user has also set the non-compliance option for the calibration reminder, the
instrument flashes once and beeps once every 10/30/60 seconds (the interval is
set by the user, and this beep function can be enabled/disabled).
Calibration is required (and indicated by this icon) if:
The defined period between calibrations has been exceeded.
The sensor has failed a previous calibration.
When a sensor fails, has been removed, this icon is shown on the display, and
there are is one beep, one flash, and one vibration per second. The display shows
“- - -“.
The Peak Reading icon is displayed when a peak gas reading has been detected
in the past 24 hours. The icon is otherwise hidden.