6.4. Bump Status
Note: When forced bump and calibration are enabled, the instrument will power down if it is not
bump tested or calibrated.
This icon is shown when less than 12 hours remains before a bump test should be
performed. The instrument flashes once and beeps once every 10/30/60 seconds, as set
by the user, when the user has also set the non-compliance option for the bump test
reminder.
The icon blinks twice per second if there is a failed or skipped bump test, and the instrument beeps
three times every 60 seconds.
A bump test is required (and indicated by this icon) if:
The defined period between bump tests has been exceeded (calibration overdue).
The sensor has failed a previous bump test.
Failed bump test notification is shown on the screen if a bump test is failed. In the information
menu, it shows that the calibration must be performed “NOW.”
Note: the sensor should be challenged on a regular basis (before each day’s use) to ensure proper
sensor performance.
6.5. Sensor Fault Status
When a sensor has been removed or failed, this icon is shown on the display, and there is one
flash, one beep, and a vibration alert every second. This display shows “- - -“.