9.9.1.2. Force Bump
When this feature is enabled, the display readsBump Now. The user cannot pass this screen
without bump testing either manually or through IntelliDoX before proceeding to Idle Screen.
For Manual Bump Testing: Press and hold the button to start the bump process. Follow the
instructions for the bump process. Once the bump test has passed, the user may proceed to the
idle screen. If the bump test fails, then the instrument will revert to the Bump Nowscreen. If no
action is taken, then the unit will automatically shut off after 60 seconds.
The default setting for bump and calibration reminders is off.
9.10. Data Logging
Turn datalogging on or off and set the frequency interval (in seconds) for data collection.
Note: Turning data logging off hides the Interval setting option. The default setting for Intervalis
005 seconds, and the interval can be set between 001 and 300 seconds.
The default setting for data logging is on.
9.11. BLE
On instruments that feature BLE (Bluetooth Low Energy), turn BLE communication on or off, so that
the instrument is discoverable or not discoverable.
Note: If BLE operation is desired, this option must be specified at time of order.
When BLE is turned off, the BLE “off” icon is displayed.
When BLE is turned on, the BLE “on” icon is displayed.
The default setting for BLE is on.
9.12. Assigned
The instrument can be assigned to a person, and their name can be programmed into it. You can
also turn assignment on or off, or edit/rename an assignment. The name can be up to 10
characters long.
The following are available for entry:
Digits 0 to 9
Upper-case characters A to Z
A space
9.13. Location
A location’s name can be programmed. You can turn location on or off, and name, edit, or rename it.
The following are available for entry:
Digits 0 to 9
Upper-case characters A to Z
A space