17. Sensors & Settings
The available sensors and their range, resolution, temperature range, and low and high setpoints
are shown here:
Gas
Type
Range
Resolution
Temperature
Low
Setpoint
High
Setpoint
H
2
S*
0-200 ppm
0.1
-40 to +6 C/-40 to 140°F
10 ppm
15 ppm
CO*
0-2000 ppm
1 ppm
-40 to +6 C/-40 to 140°F
35 ppm
200 ppm
O
2
*
0-30% v/v
0.1 %v/v
-40 to +6 C/-40 to 140°F
19.5 %v/v
23.5 %v/v
CO
2
*
0-50000 ppm
100 ppm
-20 to +60° C/-4 to 140°F
5000 ppm
30000 ppm
H
2
S
(extended
range)
0-500 ppm
0.1*
-40 to +5 C/-40 to 122°F
10 ppm
15 ppm
CO-H
0-2000 ppm
0.5 ppm
-30 to +5 C/-22 to 122°F
35 ppm
200 ppm
NH
3
0-100 ppm
1 ppm
-20 to +4 C/-4 to 104°F
25 ppm
50 ppm
NH
3
(extended
range)
0-1000 ppm
1 ppm
-20 to +40° C/-4 to 104°F
25 ppm
50 ppm
SO
2
0-100 ppm
0.1 ppm
-40 to +50° C/-40 to 122°F
2 ppm
5 ppm
HCN
0-100 ppm
0.1 ppm
-20 to +50° C/-4 to 122°F
4.7 ppm
10 ppm
Cl
2
0-50 ppm
0.1 ppm
-20 to +40° C/-4 to 104°F
0.5 ppm
1.0 ppm
NO
0-250 ppm
0.2 ppm
-30 to +50° C/-22 to 122°F
25 ppm
25 ppm
NO
2
0-100 ppm
0.1 ppm
-20 to +50° C/-4 to 122°F
2 ppm
5 ppm
PH
3
0-5 ppm
0.1 ppm
-20 to +50° C/-4 to 122°F
0.3ppm
1.0 ppm
ETO
0-100 ppm
0.1 ppm
-30 to +50° C/-22 to 122°F
1 ppm
5 ppm
ClO
2
0-1 ppm
0.01 ppm
-20 to +40° C/-4 to 104°F
0.10 ppm
0.30 ppm
O
3
0-1 ppm
0.01 ppm
-40 to +50° C/-40 to 122°F
0.10 ppm
0.20 ppm
H
2
0-1000 ppm
2 ppm
-20 to +50° C/-4 to 122°F
100 ppm
500 ppm
* 1-Series sensor