Install Instructions

H7625B, H7626B, H7635B,C H7636B, H7655B, H7656B
62-0330—01 6
APPENDIX
RH Test and Configuration DIP Switch Settings (Table 3)
IMPORTANT
Only adjust these switches for troubleshooting or
recalibrating the sensor.
(Adjustment is not normally necessary.)
Test mode will make the transmitter output a fixed 0%,
50%, or 100% value. The sensor will not affect the
transmitter output. This is used for troubleshooting or
testing only.
Switches 1, 2, and 3 are used for test mode. The output
will be a fixed 0%, 50%, or 100% signal that corresponds
to the output selected with switches 6, 7, and 8. Refer to
Figure 4 for switch settings.
Fig. 8. Settings
Table 3. Test and Calibration Settings (8-Switch Block).
* = Switch setting does not affect output
** = This switch needs to be activated first
— = Indicates OFF
Setting 12345678
4-20mA Output ————————
0-5 VDC Output —————OnOn
0-10 VDC Output ——————OnOn
0% RH Output On———————
50% RH Output —On——————
100% RH Output ——On—————
Increment RH Output ———On**On———
Decrement RH Output ———On**—On——
Reset to Original Calibration ———On**OnOn——
Reverse or Direct Acting On On** On
Reverse or Direct Acting Reset On On** On On
1
ON
2345678
WIRE CONNECTIONS
FACTORY
USE ONLY
PUSH TAB IN AND PULL ON THE SENSOR
CABLE TO REMOVE FROM J13 CONNECTOR
RED
FACTORY
USE ONLY
M31322
SW1
TB1
J2
J13
J14
4-20mA VIN VOUTCOM
P1P2
ZEROSPAN
GREEN
BLUE
1
ON
2345678
TEST SELECTION
SWITCHES (SW1)
0% RH OUTPUT
ON
OFF
1
ON
2345678
ON
50% RH OUTPUT
OFF
1
ON
2345678
ON
100% RH OUTPUT
OFF
SW1
SW1
SW1