User`s guide
51 
Scale Calibration Procedure (lbs. or kg) with Remote Display* 
4.  Reinstall the platter onto the scanner/Diva scale.  
Figure 44. Install Platter 
5.  Go to Group 1 Step 8 (Weight Unit) of the Service Mode by pressing the Zero ( ) key several times until 
‘Group 1 Step 8’ is displayed on the Remote Scale Display screen (see Figure 45 below). 
x  Depending on the current Weight Unit setting, this value will either be 1, 2, 3, or 4. 
Figure 45. Remote Scale Display Group 1 Step 8 example shown 
6.  Verify the weight unit setting. Verify the Group 1 Step 8 setting using the table shown below. Press the 
Zero ( ) key if correct. If the lb (Pounds) or kg (Kilograms) basic range setting needs to be changed, 
press the Tare ( ) key to scroll through the selectable options. Press the Zero (  ) key to accept the 
selected weight unit range. 
Group 1 Step 8 (x) 
Selectable Weight Unit Ranges 
1  lbs range (0 to 30 lbs) ** 
2  kg range (0 to 15kg) 
3  kg multi-interval (0 to 6kg, 6 to 15kg) ** 
4  lbs multi-interval (0 to 15 lbs, 15 to 30 lbs) 
** Common ranges used as default setting. 
*These procedures are for MS2420/MS2430 Scanner/Diva Scale models only. 