user manual

Configuration
90 UDA2182 Universal Dual Analyzer Product Manual January 2009
Accessing Auxiliary Menu
Press
Setup
to display the Main menu.
Use the
keys to select “Auxiliary” then press
Enter
to enter the sub-menu.
Press
to highlight the desired menu selection then press
Enter
to display the
group of parameters.
Refer to “Section 6.4.1 – “General Rules for Editing”.
Table 6-12 Auxiliary Configuration
Sub-menu
selection
Parameter Selection or Range of Setting Parameter Definition
In A Source
In B Source
Any Analog Signal
See Table 6-3
Analog Signal Source – Process signal
to be monitored by the Alarm. Any
analog source such as PV, Temperature,
Pharma, Math, Function Generator,
Switch, PID, or Calculated Values*
* units of measure between the two input
boards must be similar
Switch 1
Switch 2
Select B
Any Digital Signal
See Table 6-4
Digital Signal Source when active will
select the B input source of the switch as
the output
Switch
InA
InB
Select B
If Select B is OFF then Switch Output = In A
If Select B is ON then Switch Output = In B
Source
None
Input 1 PV
Input 2 PV
PID 1 Out
PID 2 Out
Function Generator selections have 2
input sources
(Input 1 PV and Input 2 PV).
Func Gen 1
Func Gen 2
ATTENTION
The X (n) value must be < X(n+1) value. Thus, if fewer than 11 breakpoints are needed,
be sure to configure any unneeded breakpoints with the same X and Y values used for the
previous breakpoint.