6.15.3 Auto Cycling Configuration
Table 6-16 Auto Cycling Configuration
Sub-menu
selection
Parameter Selection or Range
of Setting
Parameter Definition and Restrictions
Auto Cycling
Disable (default)
Enable
Allows auto cycling to be selected. This
should be enabled after configuration is
complete.
Hold Active
Enable (default)
Disable
When enabled, the output(s) sourced by input
n for Auto Clean n is in hold during auto
cycling.
Probe Transit
Disable (default)
Enable
When enabled, allows probe extract and
probe insert sequence steps to occur and
automated probe extract and insert
parameters are made available.
Cycle Start Source
None or Any Digital
Signal
See Table 6-4
Starts Auto Cycle on specific Digital Signal
selected changing from 0 to 1.
Extract Wait Src
(Probe Transit =
Enabled)
None or Any Digital
Signal
See Table 6-4
Allows selection of a specific Digital Signal
that causes a delay in the probe extraction
sequence. While the selected digital input is
active the probe extraction will not end unless
a timeout occurs as determined by the
duration configured in Probe Transit Mins. If
a source is configured, the timeout results in
an Auto Cycle Fail. If a digital signal is not
available, the source may be left at “None”
and the extract step will occur for the duration
of Probe Transit Mins.
Insert Wait Src
(Probe Transit =
Enabled)
None or Any Digital
Signal
See Table 6-4
Allows selection of a specific Digital Signal
that causes a delay in the probe insertion
sequence. While the selective digital input is
active the probe insertion will not end unless a
timeout occurs as determined by the duration
configured in Probe Transit Mins. If a source
is configured, the timeout results in an Auto
Cycle Fail. If a digital signal is not available,
the source may be left at “None” and the
extract step will occur for the duration of
Probe Transit Mins.
Cycle Interval
Off(default)
Monthly
Weekly
Daily
Custom
Frequency of Auto Cycle occurrence
Auto Cycle 1
Auto Cycle 2
Start Time
(Custom)
Disable (default)
Enable
Set specific time for Auto-Cycle to start.