user manual

Configuration
102 UDA2182 Universal Dual Analyzer Product Manual January 2009
Sub-menu
selection
Parameter Selection or Range
of Setting
Parameter Definition and Restrictions
Start Day
(Dependent
parameters)
1 to 28 (default = 1)
Sunday – Saturday
(default = Sunday)
1 to 31 (default = 1)
Cycle Interval is Monthly
Cycle Interval is Weekly
Cycle Interval is Custom, Start Time enabled
Start Hour
(Dependent
parameters)
0 to 23
(default = 12)
Cycle Interval is Monthly, Weekly, Daily
Start Mins
(Dependent
parameters)
0 to 59 (default = 0)
Cycle Interval is Monthly, Weekly or Daily
Cycle Interval is Custom, Start Time enabled
Period Days
(Custom)
0 to 100(default =
0)
The period day parameter allows the selection
of how often the Auto cycle will occur.
For Example: 20 means that the Auto Cycle
will occur every 20 days.
Period Hours
(custom)
0 to 23 (default = 1)
For Example: 4 means that the Auto-cycle will
occur every 4 hours when the days and
minutes are set to 0.
Period Mins
0 to 59 (default = 0)
For Example: 30 means that the Auto-cycle
will occur every 30 minutes when the days
and hours are set to 0.
Rinse Cycle Cnt
0 to 100(default =
1)
Allows selection for frequency of rinse occurrence.
Cal Cycle Cnt
(PV is DO)
0 to 100(default =
1)
Allows selection for frequency of calibration
occurrence.
For Example: 1 indicates that a
calibration will occur every cycle while a 10
indicates that a calibration will occur every 10
th
cycle.
Cal 1 Cycle Cnt
(PV is pH)
0 to 100(default =
1)
Allows selection for frequency of calibration
occurrences.
Cal 2 Cycle Cnt
(PV is pH)
0 to 100(default =
1)
Allows selection for frequency of calibration
occurrences.
Max Transit Mins
(Probe Transit =
Enabled)
0 to 30.00
(default = 0.50)
Maximum Probe Transit time in minutes.
Rinse Mins
0 to 30.00
(default = 0.50)
Duration of Rinse sequence in minutes
Max Cal Mins
(PV pH, DO)
0 to 30.00
(default = 0.50)
Maximum calibration time in minutes. Enough time
should be entered to allow process stabilization
and a 20 second measurement time.