Configuration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 109
Sub-menu
selection
Parameter Selection or Range of
Setting
Parameter Definition
Password
0000 (default) to 9999
AAAA to ZZZZ
Setup configuration, calibration and maintenance
functions can be password-protected. The
password can be any number between 1 and
9999 or letters. (When the password is zero, the
operator will not be prompted to enter a
password.)
Follow the “ General Rules for Editing” to change
the digits.
Unit ID
Unit Identification
Read Only
Option ID*
Option ID Number
= 0 if PID is not available
= ID Number if PID available
*See note at end of table
PID Control* 1 Loop
2 Loops (default)
Enables the PID Control configuration
Parameters
*See note at end of table
Unit Reset
No (default)
Yes
Unit Reset initializes all calibration and
configuration data to factory default values, with
the exception of the Factory Temperature
Calibration correction values
Inputs
Input 1 Type
Input 2 Type
Read Only pH/ORP - pH or Oxidation Reduction Potential
pH Preamp – pH with preamplifier
Conductivity (Dual if both inputs same)
DO ppm - Dissolved Oxygen
DO ppb - Dissolved Oxygen