Input Calibration
160 UDA2182 Universal Dual Analyzer Product Manual January 2009
Table 8-8 Conductivity of Potassium Chloride Solutions at 25 °C
* M = Molarity; 1M = 74.555g potassium chloride per liter of solution
Procedure
Table 8-9 Procedure for Performing Calibration Trim Using a Reference Solution
Step Action Screen
1
Press
Calibrate
CALIBRATION
Input Temp Cal
Output Cal
Cal History
Input PV Cal
CALIBRATION
Input Temp Cal
Output Cal
Cal History
Input PV Cal
2
Press
Enter
PV INPUT CAL
In 1 pH/ORP Cal
In 2 Conduc Cal
PV INPUT CAL
In 1 pH/ORP Cal
In 2 Conduc Cal
Use
to select
Input 1 or 2 Conduc Cal
3
Press
Enter
IN2 Conduc Cal
Sample Cal
Cal Trim
Reset Cal Trim
1.00
IN2 Conduc Cal
Sample Cal
Cal Trim
Reset Cal Trim
1.00
4
Remove the cell from the
process.
Rinse the cell thoroughly
with de-ionized water.
Concentration M* Conductivity (microSiemens
per cm)
0.001 147.0
0.005 717.8
0.01 1,413
0.02 2,767
0.05 6,668