user manual

Outputs Calibration
184 UDA2182 Universal Dual Analyzer Product Manual January 2009
Viewing and resetting 20mA and 4mA Offset
If the calibration is suspect, you can reset the 20mAand 4mA Offset and calibrate again.
In the same screen as “
20mA and 4mA Offset”, use the keys to highlight “Reset
20mA Offset
” or “Reset 4mA Offset”.
OUTPUT 1
20mA Offset
4mA Offset
-147
3
Reset 20mA Offs
Reset 4mA Offs
OUTPUT 1
20mA Offset
4mA Offset
-147
3
Reset 20mA Offs
Reset 4mA Offs
Figure 9-1 Resetting Output 1 Offsets (example)
Press ENTER. The 20mA Offset or 4mA Offset will be reset to 0(default).