user manual

Operating the Analyzer
44 UDA2182 Universal Dual Analyzer Product Manual January 2009
Status
Display
Parameter Status
(Read Only)
Status Definition
WebPgLngSet EE/RT/PC Identifies the web page language set programmed into
the Communications card.
EE web pages support English, French, German,
Italian and Spanish.
RT web pages support English, Russian and Turkish
PC web pages support English, Polish and Czech
Address Value
Address states the Modbus RTU slave ID
Baud Rate Value
Baud Rate set for RS485
Word Swap Yes/No
Word Swap indicates whether the word order set for
Modbus Communications is Little Endian (NO) or Big
Endian (YES)
MACaddr Hi
MACaddr Low
Value
MACaddr Hi and MACaddr Low is the MAC address
of the Communication card
DHCP Yes/No
DHCP indicates whether Dynamic Host Configuration
Protocol is used.
DHCP is a protocol used by network devices (clients)
to obtain various parameters necessary for the clients
to operate in an Internet Protocol (IP) network.
IPaddr Value
IPaddr gives the IP address of the Communication
card.
SubnetMsk Value
SubnetMsk indicates the Subnet mask used by the
Communication card.
Gateway Value
Gateway Indicates the default Gateway IP address
used by the Communication card.
DnsSrvr Value
DnsSrvr displays the DNS (Domain Name Service)
server IP address used by the Communication card
SMTPsrvr Value
SMTPsrvr displays the SMTP (Simple Mail Transfer
Protocol) server IP address used by the
Communication card.
Auto Cycling
(if
configured)
Next Rinse 1
Next Rinse 2
Next Cal 1
Next Cal 2
Date and Time Read Only – The Status Displays menu includes an
Auto Cycling selection when any auto cycle is enabled
and its cycle timer is also enabled. This display
provides information on the next occurrences of rinses
and calibrations for any auto cycle with timer
according to the configured cycle count of each
operation.