64 UDA2182 Universal Dual Analyzer Product Manual January 2009
Sub-menu
selection
Parameter Selection or
Range of
Setting
Parameter Definition
Temp Deg F or C
(Temp Type =
Manual)
14.0 to 230.0ºF
default = 77ºF
-10 to 110ºC
default = 25ºC
Temp Deg F or C will appear depending on what
Temperature Unit was selected in “Maintenance”
setup group, parameter “Temp Units”.
Solu Temp Comp
(Not ORP)
None (default)
Custom
H
2
0
NH
3
Phosphate
Morpholine
Enter “Solution pH/ºC” value
Pure Water
Ammonia
Phosphate
Morpholine
Solution pH/ºC
(Solu Temp
Comp = Custom)
(Not ORP)
0.000 (default)
to
-0.050
Measured pH is displayed and transmitted normalized
to a solution temperature of 25°C as determined by
the current Solution Temperature Coefficient. This is
expressed in units of pH/°C with precision to the
hundredths decimal place. The parameter “Solu Temp
Coeff” allows the selection of the following entries.
Follow the “General Rules for Editing” in section
6.4.1 to make the changes. (-) Will appear when first
digit to the right of decimal point is changed.
Solution Type Temp Coefficient
None (Default)
0.000
H
2
O (Pure Water)
-0.016
NH
3
(Ammonia)
-0.032
PO
4
(Phosphate)
-0.032
C
4
H
9
NO (Morpholine)
-0.032
Custom
User Entry
PV Bias -99999 to
99999
default = 0.00
PV Bias Constant - is used to compensate the input
for drift of an input value.
Failsafe -99999 to
99999
default = 14.00
The output value to which the output will go to protect
against the effects of failure of the equipment.
Filter Time 0 to 120
default = 0
A software digital filter is provided for dampening the
process noise. This filter is applied before the limit
functions.