70 UDA2182 Universal Dual Analyzer Product Manual January 2009
Sub-menu
selection
Parameter Selection or
Range of
Setting
Parameter Definition
Solu Temp Comp None
Custom
H20
NH3
PO4
C4H9NO
HCl
NaCl (default)
H2SO4
NaOH
Measured Conductivity and Resistivity can optionally
be temperature compensated to 25°C for a specific
solution type. TDS and concentration are always
measured based on a specific solution type. The cell
constant and measurement type determines which
solution types are available for selection, according to
the table below:
Cell
Constant
Available Selectable Solution Types
Use the keys to select
0.01 None (Conductivity/Resistivity only),
NaCl (
μS/cm, mS/cm, TDS ppb, TDS ppm ),
NH
3
(μS/cm, TDS ppb, TDS ppm ),
C
4
H
9
NO (μS/cm, TDS ppb, TDS ppm ),
H2SO4;HCL;NaOH (mS/cm)
0.1
(Default)
None (Conductivity/Resistivity only),
NaCl(
μS/cm, mS/cm, TDS ppb, TDS ppm ),
NH
3
(μS/cm, TDS ppb, TDS ppm ),
C
4
H
9
NO (μS/cm, TDS ppb, TDS ppm ),
H
2
SO
4
;HCL;NaOH (mS/cm)
1 None (Conductivity only),
NaCl (
μS/cm, mS/cm, TDS ppm, TDS ppt ),
NH
3
(μS/cm, TDS ppm ),
C
4
H
9
NO (μS/cm, TDS ppm ),
H
2
SO
4
;HCL;NaOH (mS/cm, TDS ppt)
10 None (Conductivity only),
NaCl (
μS/cm, mS/cm, TDS ppm, TDS ppt ),
NH
3
(μS/cm, TDS ppm ),
C
4
H
9
NO (μS/cm, TDS ppm ),
H
2
SO
4
;HCL;NaOH (mS/cm, TDS ppt)
25 None (Conductivity only),
HCl (mS/cm, Concentration),
NaCl (
μS/cm, mS/cm, Concentration),
H
2
SO
4
(mS/cm, Concentration),
NaOH (mS/cm, Concentration)
50 None (Conductivity only),
HCl (mS/cm, Concentration),
NaCl (
μS/cm, mS/cm, Concentration),
H
2
SO
4
(mS/cm, Concentration),
NaOH (mS/cm, Concentration)
Wire Len Feet
+
0 to 1000 ft
default = 0
Wire Len Meters
+
0 to 304.80
default = 0
Refer to appendix 15.2 to enter values for lead wire
resistance compensation