Configuration
January 2009 UDA2182 Universal Dual Analyzer Product Manual 71
Sub-menu
selection
Parameter Selection or
Range of
Setting
Parameter Definition
Wire Size AWG
+
16 AWG
18
AWG(default)
20
AWG
22
AWG
Wire Size Sq mm
+
0.33 to 2.08
default = 0.82
Pharma Type None
PhEur
USP
default = None
PhEur - Pharmacopoeia Europa
USP - United States Pharmacopoeia
standard procedure stages for determining Purified
Water
Pharma PV High -99999.00 to
99999.00
(default
10.000)
Pharma PV High Value – Measured solution
conductivity value scaled for 100%
Pharma PV Low -99999.00 to
99999.00
(default 0.000)
Pharma PV Low Value - Measured solution
conductivity value scaled for 0%
Pharm Tmr Mins 000.0 to 120.0
(default
10.000)
Pharma Timer Minutes - If the Pharma sample does
not pass the Stage 1 conductivity requirement a Fail
signal is generated, then the State 2 and Stage 3
tests are conducted. When the Stage 2 or Stage 3 test
is successful, the fail signal is cancelled and the
Pharma Timer begins to count down from the
configured minutes value set here. When the Timer
countdown is completed, the Pharma function block
returns to Stage 1.
PV Bias -9999.00 to
9999.00
default = 0.000
PV Bias Constant - is used to compensate the input
for drift of an input value.
Failsafe 0.0 to 2000
default =
2000.000
The output value to which the output will go to protect
against the effects of failure of the equipment.
Filter Time 0 to 120.0
default = 0.000
A software digital filter is provided for dampening the
process noise and is applied before the limit functions.