6.2.5 XNX EC Sensor Cross-sensitivity
Gas type Part Number Gas Type Applied Concentration Unit Reading Unit
O
2
XNXXS01SS
XNXXSO1FM
Carbon Dioxide 5 %vol 0.1
%vol (change O
2
reading) per %vol CO
2
HCl XNXXSR1SS
Carbon monoxide 2000
ppm
0
ppm HCl
Hydrogen 20000 0
Chlorine 5 5.6
Nitrogen dioxide 5 0.9
Propan-2-ol 500 0
Methanol 500 0
Hydrogen fluoride 5 6.7
Hydrogen suflfide 25 -3.6
Sulphur dioxide 50 22.4
Arsine 1 0
Phosphine 1 -0.14
Diborane 1 -1.3
H
2
S
(Low Range)
XNXXSH3SS
Ammonia 50
ppm
0 ppm H
2
S
Carbon Monoxide 100 <2 ppm H
2
S
Carbon Dioxide 5000 0 ppm H
2
S
Chlorine 0.5 0 ppm H
2
S
Ethylene 100 0 ppm H
2
S
Hydrogen 100 0 ppm H
2
S
Hydrogen Sulfide 10 10 ppm H
2
S
Nitrogen Monoxide 25 0 ppm H
2
S
Nitrogen Dioxide 3 0 ppm H
2
S
Sulfur Dioxide 2 0 ppm H
2