GasAlertMaxXT II
Display Elements
Table 4. Pushbutton
Pushbutton Description
C
To activate the detector press C.
To deactivate the detector, press and hold C until the OFF countdown is complete and the LCD deactivates.
To view the date/time, TWA, STEL, and maximum (MAX) readings, press C twice rapidly. To clear the TWA,
STEL, and MAX readings, press C when the LCD displays CLEAR ALL.
To initiate calibration, press and hold C as the detector performs the OFF countdown. Continue holding C
while the LCD briefly deactivates and then reactivates to begin the CAL countdown. Release C when the
CAL countdown is complete.
To activate the backlight, press C and release.
To acknowledge the latched and pump alarms, press C.
To acknowledge a low alarm and disable the audible alarm temporarily, press C. The Low Alarm
Acknowledge option must be enabled in Fleet Manager II.
To acknowledge any of the Due Today alarms (calibration, bump test, pump block test), press C.