GasAlertMaxXT II
Startup Tests
12. The LCD next displays the number of days remaining
until the next bump test is due.
Force Bump Enabled (optional)
If the Force Bump is enabled in Fleet Manager II and a
sensor is overdue for a bump test, the sensor(s) must be
tested to continue and enter normal operation. The
following screen displays.
Apply gas while BUMP DUE NOW displays. Verify the
visual, audible, and vibrator alarms activate. The detector
must enter alarm to pass the bump test. For complete
instructions, refer to Bump Test
.
If Force Bump is disabled, press C to acknowledge the
warning and continue with the startup tests.
Note
If the bump test interval field is set to 0 in Fleet Manager II, the
startup bump test is bypassed. Refer to Bump Interval
in the
Device Configuration
.
The detector remains in alarm until the gas dissipates and
clears from the sensors. Verify that the visual and audible
alarms activate.
Startup Test Pass
If the detector passes all of the startup tests, the detector enters normal
operation and displays the ambient gas readings.
The detector begins recording the maximum gas exposure (MAX), and
calculating the short-term exposure level (STEL) and time-weighted
average (TWA) exposures.