Product Manual

Table Of Contents
23
GasAlertMaxXT II
Device Configuration
Refer to the following options for descriptions and functionality.
Note
The Serial Number, Firmware Version, and Hardware
Version fields require no data entry. Settings for these fields
are factory defined.
Serial Number Field
This field displays the serial number (e.g. MA110-001000) of the
detector.
Firmware Version
This field displays the current firmware version (e.g. 02_000) that
displays on the detector LCD during the startup tests. If new firmware
is uploaded to the detector, the Firmware Version field automatically
updates.
Hardware Version
This field displays the current version of hardware the detector is
operating with.
Startup Message Top Line
Enter a line of text that will display on the detector LCD during startup
(maximum 25 characters, spaces included). Enter any type of informa-
tion such as employee name, plant, area, emergency number(s), etc.
Depending upon the length of the message, text will either display or
scroll across the top line of the LCD.
Startup Message Bottom Line
Enter a line of text that will display on the detector LCD during startup
(maximum 25 characters, spaces included). Enter any type of informa-
tion such as employee name, plant, area, emergency number(s), etc.
Depending upon the length of the message, text will either display or
scroll across the bottom line of the LCD.
The user options define the settings for the detector.
Note
When options are enabled/disabled, the checkbox displays with a
red frame until the new settings are saved to the detector.
Lockout on Self-Test Error (failed sensor lock)
If enabled and a sensor fails during startup, the following screens display
and the detector deactivates.
To enter normal operation, the sensor must be operating correctly.
Refer to Troubleshooting
and Replacing a Sensor or Sensor Filter.
The detector is shipped with the Lockout on Self-Test Error option
disabled.