GasAlertMaxXT II
Troubleshooting
Table 7. Troubleshooting
Problem Possible Cause Solution
Detector does not display normal
ambient gas reading after startup tests.
Sensors not stabilized Used sensor: Wait 60 seconds
New sensor: Wait 5 minutes
Sensor(s) requires calibration Calibrate the sensor(s). Refer to Calibration
.
Target gas is present Detector is operating properly. Use caution in
suspect areas.
Detector does not respond to
pushbutton.
Battery is depleted Charge the battery. Refer to Charging the Bat-
tery.
Detector is performing operations that do
not require user input
Pushbutton operation restores automatically
when the operation ends.
Detector does not accurately
measure gas.
Sensor(s) requires calibration Calibrate the sensor(s). Refer to Calibration
.
Detector is colder/hotter than gas
temperature
Allow the detector to attain ambient
temperature before use.
Sensor filter is blocked Replace the sensor filter. Refer to Replacing a
Sensor or Sensor Filter.
Detector does not enter alarm. Alarm setpoint(s) are set incorrectly Reset the alarm setpoints. Refer to Resetting
Gas Alarm Setpoints.
Alarm setpoint(s) set to zero Reset the alarm setpoints. Refer to Resetting
Gas Alarm Setpoints.
Detector is in calibration mode Complete the calibration procedure.