Table 7. Troubleshooting
Problem Possible Cause Solution
Detector intermittently enters alarm
without reason.
Ambient gas levels are near alarm
setpoint or the sensor is exposed to a
puff of the target gas
Detector is operating normally. Use caution in
suspect areas. Check maximum gas exposure
reading.
Alarms are set incorrectly Reset the alarm setpoints. Refer to Resetting
Gas Alarm Setpoints.
Sensor(s) requires calibration Calibrate the sensor(s). Refer to Calibration
.
Missing or faulty sensor(s) Replace the sensor(s). Refer to Resetting Gas
Alarm Setpoints.
Battery has been charging for 6 hours.
Charging indicator on LCD shows the
battery is still charging.
Battery is trickle charging Battery is fully charged and is ready for
operation.
Battery indicator does not display when
charging.
Battery is depleted below normal
levels
Charge the battery for 8 hours. If the battery
indicator does not light after charging, contact
BW Technologies by Honeywell
.
Pump is not operating. Diffusion cap is off or not attached
correctly.
Attach diffusion cap. Ensure seal is tight.
There is a blockage in the tubing. Clear the blockage.
Particulate and/or moisture filter requires
replacement
Replace the filter(s). Replacing the Pump Fil-
ters.
Features and options not operating as
expected.
Changes in Fleet Manager II Verify that the settings in Fleet Manager II are
correct.