5 - 22 Select Mode
430-04685-001 (Rev. 7/05)
PROGRAMMING THE 4 (OR 1) HOUR DOSE LIMIT
The opportunity to program this feature is presented in two entry
screens. The first screen prompts the clinician to choose whether
or not to set a Dose Limit. Selecting N
O> brings up the CONFIRM NO
DOSE LIMIT SCREEN. Selecting YES> brings up the DOSE LIMIT ENTER
VALUE SCREEN. This requires that the clinician enter a value within
the displayed range.
NOTE: If Hospira MedNet
TM
Software is enabled, and a
upper hard or soft limit for that CCA and drug has been
assigned, then a
DOSE LIMIT entry is required.
PROGRAMMING A DOSE LIMIT
1. Select YES> at the DOSE LIMIT
SELECTION SCREEN.
The display advances to the
DOSE LIMIT
ENTER VALUE SCREEN.
2. Enter the 4 HOUR DOSE LIMIT value
using the numeric keypad.
NOTE: Entry of zero for the 4 HR
Dose Limit, using the numeric
keypad, is NOT accepted by the
PCA Infuser; see
Programming No
Dose Limit on page 5 - 24
.
3. Then press .
MORPHINE PF
1 mg/mL
SET A DOSE
LIMIT?
YES
NO
PREVIOUS
MORPHINE PF
1 mg/mL
ENTER 4 HOUR
DOSE LIMIT
THEN PRESS
ENTER BUTTON
.1-80 mg
40 mg
milligrams
PREVIOUS
ENTER