User's Manual
Table Of Contents
- System Operating Manual ®
- Table of Contents
- Section 1 Descriptive Information
- Section 2 Principles of Operation
- Section 3 Equipment Description
- Section 4 Basic Operation
- Section 5 Select Mode
- 5.1 Modes of Delivery
- 5.2 PCA Only Mode
- 5.3 Continuous Mode
- 5.4 PCA+Continuous Mode
- 5.5 Protocols
- 5.6 Dose Limit
- 5.7 Clearing the History and/or Rx Settings
- 5.8 Using Review/Confirmation Screens
- 5.9 Changing Settings During Setup
- 5.10 Making Changes After Setup
- 5.11 Clearing Shift Totals
- 5.12 Changing a Vial
- 5.13 Adding a Supplemental Loading Dose
- 5.14 Checking History & Settings
- 5.15 History and Event Log
- Section 6 Advanced Programming with Hospira MedNetTM Software
- Section 7 Troubleshooting
- Section 8 Maintenance
- Section 9 Specifications
- Section 10 Prescription Delivery Limits
- Appendix A Generating Bar Code Labels (Hospira MedNetTM Software Enabled)
- Appendix B Network Printer Setup
- Appendix C Warranty
- Appendix D Contact Information
- Index
1- 14 Descriptive Information
430-04685-001 (Rev. 7/05)
PROGRAMMING
This section presents known infuser programming cautions.
WARNING
For custom syringes, confirm that the displayed
concentration (mg/mL) or (mcg/mL) exactly matches
the concentration value and drug name on the syringe.
If they do not match, under/overdosage may result.
• In the CONTINUOUS and PCA+CONTINUOUS modes, if a
purge is not performed after a syringe change, the infuser auto-
matically performs a small system compliance step to remove
slack when the [START/PAUSE] key is pressed (with the door
locked). Although, fluid is not normally delivered to the patient
during the compliance step, under some conditions up to 0.3
mL of fluid may be delivered. If 0.3 mL of fluid represents a haz-
ard to the patient, disconnect the set during this operation.
• At flow rates less than 0.5 mL/hr, there may be a significant
delay before flow is established if the system is not purged.
• Selections are rounded up to the nearest tenth of a digit for mg/
mL values or to the nearest digit for mcg/mL values.
LOADING DOSE/DOSE LIMITS
This section presents Loading Dose and Loading Dose
information and cautions.
• The loading dose is always included in the total dose delivered.
• Setting a new dose limit will not erase the previous dose history.
• Always monitor the PCA Infuser when delivering medication
with the door open.
• Patient Pendant is only to be pressed by the intended patient.
WARNING
Patient Pendant is only to be pressed by the patient.