User's Manual

Table Of Contents
LifeCare PCA Infusion System 5 - 11
430-04685-001 (Rev. 7/05)
12. Enter a continuous rate using the
keypad. The value range is
displayed on the screen.
If value is entered incorrectly, press
to change value.
13. Then press .
If programming changes need to be
made, select P
REVIOUS> to return to the
preceding screen.
14. Set a specific dose limit by
selecting
YES> and advancing to
the Dose Limit Enter Value
SCREEN. Select NO> to choose No
Dose Limit.
No Dose Limit will be selected for this
example.
If Hospira MedNet
TM
Software is
enabled and a upper hard or soft limit
for that CCA and drug, then a
DOSE
LIMIT entry is required, and this screen
will not be displayed.
NOTE: The pre-programmed Dose Limit is 1 or 4 hours.
The Hospira MedNet
TM
Software range is larger (1, 4, 6,
or 12 hours).
For complete information on this feature,
see Dose Limit on page 5 - 21
.
ENTER
CONT. RATE
THEN PRESS
ENTER BUTTON
PREVIOUS
m
illigrams
0.5-100 mg/hr
2
mg
hr
MORPHINE PF
1 mg/mL
C
L
R
ENTER
SET A DOSE
LIMIT?
YES
NO
PREVIOUS
MORPHINE PF
1 mg/mL