User's Manual
Table Of Contents
- System Operating Manual ®
- Table of Contents
- Section 1 Descriptive Information
- Section 2 Principles of Operation
- Section 3 Equipment Description
- Section 4 Basic Operation
- Section 5 Select Mode
- 5.1 Modes of Delivery
- 5.2 PCA Only Mode
- 5.3 Continuous Mode
- 5.4 PCA+Continuous Mode
- 5.5 Protocols
- 5.6 Dose Limit
- 5.7 Clearing the History and/or Rx Settings
- 5.8 Using Review/Confirmation Screens
- 5.9 Changing Settings During Setup
- 5.10 Making Changes After Setup
- 5.11 Clearing Shift Totals
- 5.12 Changing a Vial
- 5.13 Adding a Supplemental Loading Dose
- 5.14 Checking History & Settings
- 5.15 History and Event Log
- Section 6 Advanced Programming with Hospira MedNetTM Software
- Section 7 Troubleshooting
- Section 8 Maintenance
- Section 9 Specifications
- Section 10 Prescription Delivery Limits
- Appendix A Generating Bar Code Labels (Hospira MedNetTM Software Enabled)
- Appendix B Network Printer Setup
- Appendix C Warranty
- Appendix D Contact Information
- Index
LifeCare PCA Infusion System 5 - 19
430-04685-001 (Rev. 7/05)
6. If YES> is selected, disconnect the set from the patient, and
press and hold the
PURGE>.
7. Select
YES> to continue, or select NO> to purge again until
complete.
8. After purging, reconnect set to patient.
9. Set Loading Dose (if desired) by selecting
YES>. For more
information see
Loading Dose on page 4 - 31.
10. Enter a Loading Dose in the displayed range.
11. Then press .
12. Press to infuse programmed
Loading Dose. Select
DELIVER
LATER> to delay loading dose
delivery. From the
SELECT
DELIVERY MODE SCREEN, select
PROTOCOLS>.
13. Press to accept protocol,
or select
NEXT PROTOCOL> and
PREVIOUS> to view other stored
protocols if available.
ENTER
MORPHINE PF
1 mg/mL
SELECT
DELIVERY MODE
PREVIOUS
PCA ONLY
PCA + CONT.
CONTINUOUS
PROTOCOLS
LOCKOUT
4 HOUR LIMIT
2 mg
NONE
PCA DOSE
10 min
NEXT PROTOCOL
PREVIOUS
MORPHINE PF
PRESS ENTER
TO SELECT
#3
ENTER