T10F A0608-14.0 HBM: public 75
Nominal (rated) torque
M
nom
10k5k3k2k1k50020010050NVm
Impact resistance, test
severity level according to
DIN IEC 68; Part 227; IEC
682271987
Number n 1000
Duration ms 3
Acceleration (half sine) m/s
2
650
Vibration resistance, test
severity level per
DIN IEC 68, Part 2-6:
IEC 68‐2‐6‐1982
Frequency range Hz 5...65
Duration h 1.5
Acceleration (amplitude) m/s
2
50
Nominal speed (x1000) min
-1
15 15 15 12 12 10 10 8 8
Load limits
4)
Limit torque, related to
M
nom
% 400 200 160
Breaking torque, related to
M
nom
% >800 >400
>30
0
Longitudinal limit force kN 2 2 4 7 7 12 14 22 31
Lateral limit force kN 1 1 3 6 8 15 18 30 40
Bending limit moment Nm 70 70 140 500 500 1000 1600 2500 4000
Oscillation width per
DIN 50 100 (peaktopeak)
5)
kNm 0.16 0.16 0.32 0.8 1.6 3.2 4.8 8.0 12.0
1)
RS-422 complementary signals; factory settings version SF1/SU2
2)
Factory settings version KF1 (changeover not possible)
3) C
an be switched off
4)
Each type of irregular stress (bending moment, lateral or longitudinal force, exceeding nominal
(rated) torque) can only be permitted up to its specified static load limit provided none of the others
can occur at the same time. If this condition is not met, the limit values must be reduced. If 30% of
the bending limit moment and lateral limit force occur at the same time, only 40% of the
longitudinal limit force is permissible and the nominal (rated) torque must not be exceeded. The
permissible bending moments, longitudinal forces and lateral forces can affect the measurement
result by approx. 1 % of the nominal (rated) torque.
5)
It is permissible to exceed the nominal (rated) torque by 100% with T10F/50 Nm, but the nominal
torque may not be exceeded with T10F/100 Nm up to 10 kNm.