User's Manual
Table Of Contents
- T12
- Contents
- Safety instructions
- 1 Markings used
- 2 Scope of supply
- 3 Operation
- 4 Application
- 5 Signal flow
- 6 Structure and mode of operation
- 7 Mechanical installation
- 7.1 Important precautions during installation
- 7.2 Conditions on site
- 7.3 Mounting position
- 7.4 Installing the slotted disc (rotational speed measuring system only)
- 7.5 Installing the rotor
- 7.6 Fitting the protection against contact (option)
- 7.7 Installing the stator
- 7.8 Optical rotational speed/angle of rotation measuring system (option)
- 8 LED status display
- 9 Electrical connection
- 10 Shunt signal
- 11 Load‐carrying capacity
- 12 TEDS
- 13 Maintenance
- 14 Waste disposal and environmental protection
- 15 Specifications
- 16 Dimensions
- 16.1 Rotor 100 N⋅m to 200 N⋅m
- 16.2 Rotor 500 N⋅m to 10 kN⋅m
- 16.3 Stator 100 N⋅m to 200 N⋅m with rot.speed meas. system
- 16.4 Stator 100 N⋅m to 200 N⋅m with rot. speed meas. system
- 16.5 Stator 100 N⋅m to 10 kN⋅m with rot. speed meas. system
- 16.6 Stator 100 N⋅m to 200 N⋅m with prot. against contact
- 16.7 Stator 100 N⋅m to 200 N⋅m with prot. against contact
- 16.8 Stator 500 N⋅m to 1 kN⋅m with prot. against contact
- 16.9 Stator 2 kN⋅m to 10 kN⋅m with prot. against contact
- 16.10 Mounting dimensions
- 17 Supplementary technical information
- 18 Condition at the time of delivery
- 19 Ordering numbers
- 20 Accessories
59
T12
A1979−10.0 en HBM
Nominal (rated) torque M
nom
Nm 100 200 500
kNm 1
Temperature effect per 10 K in the nominal
(rated) temperature range
on the output signal, related to the actual value
of the signal span
%
< 0.03
on the zero signal % < 0.03
Residual ripple mV < 3
Angle of rotation
Accuracy degrees 1 (typ. 0.1)
Resolution degrees 0.01
Correction of runtime deviation between
torque LP1 and the angle of rotation for filter
frequencies
Hz
4000; 2000; 1000; 500; 200; 100
Measuring range degrees
0 to 360 (single-turn) to "1440
(multi-turn)
Performance
Measurement frequency range Hz 80 (−1 dB)
Resolution W 1
Full scale value W
P
max
+ M
nom
@ n
nom
@
p
30
[M
nom
] in Nm
[n
nom
] in rpm
Temperature effect per 10 K in the nominal
(rated) temperature range on the power signal,
related to the full scale value
%
"0.05@n/n
nom
Non-linearity including hysteresis, related to
the full scale value
%
"0.02@n/n
nom
Sensitivity tolerance (deviation of the actual
measurement signal span of the power signal
related to the full scale value)
%
"0.05
Temperature signal of the rotor
Accuracy K 1
Measurement frequency range Hz 5 (−1 dB)
Resolution K 0.1
Physical unit − C
Data rate
Meas.
values/
s
40