User's Manual
Table Of Contents
- T12
- Contents
- Safety instructions
- 1 Markings used
- 2 Scope of supply
- 3 Operation
- 4 Application
- 5 Signal flow
- 6 Structure and mode of operation
- 7 Mechanical installation
- 7.1 Important precautions during installation
- 7.2 Conditions on site
- 7.3 Mounting position
- 7.4 Installing the slotted disc (rotational speed measuring system only)
- 7.5 Installing the rotor
- 7.6 Fitting the protection against contact (option)
- 7.7 Installing the stator
- 7.8 Optical rotational speed/angle of rotation measuring system (option)
- 8 LED status display
- 9 Electrical connection
- 10 Shunt signal
- 11 Load‐carrying capacity
- 12 TEDS
- 13 Maintenance
- 14 Waste disposal and environmental protection
- 15 Specifications
- 16 Dimensions
- 16.1 Rotor 100 N⋅m to 200 N⋅m
- 16.2 Rotor 500 N⋅m to 10 kN⋅m
- 16.3 Stator 100 N⋅m to 200 N⋅m with rot.speed meas. system
- 16.4 Stator 100 N⋅m to 200 N⋅m with rot. speed meas. system
- 16.5 Stator 100 N⋅m to 10 kN⋅m with rot. speed meas. system
- 16.6 Stator 100 N⋅m to 200 N⋅m with prot. against contact
- 16.7 Stator 100 N⋅m to 200 N⋅m with prot. against contact
- 16.8 Stator 500 N⋅m to 1 kN⋅m with prot. against contact
- 16.9 Stator 2 kN⋅m to 10 kN⋅m with prot. against contact
- 16.10 Mounting dimensions
- 17 Supplementary technical information
- 18 Condition at the time of delivery
- 19 Ordering numbers
- 20 Accessories
41
T12
A1979−10.0 en HBM
Assignment for plug 2:
Rotational speed measuring system with reference signal
Binder 423
device plug
Top view
7
3
4
6
2
5
1
8
Plug
pin
Assignment Color
code
Sub-D
plug pin
1 Rotational speed measurement signal (pulse
string, 5 V
1)
; 0)
rd 12
2
Reference signal (1 pulse/rev., 5 V
1)
)
bu 2
3 Rotational speed measurement signal
(pulse string, 5 V
)
; phase-shifted 90) gy 15
4 Reference signal (1 pulse/rev., 5 V
1)
) bk 3
5
TEDS for rotational speed
vt 9
6 Rotational speed measurement signal (pulse
string, 5 V
1)
; 0)
wh 13
7
Rotational speed measurement signal
(pulse string, 5 V
)
; phase-shifted 90)
gn 14
8
Measurement signal 0 V
bk
2)
8
Shielding connected to housing ground
1)
RS−422 complementary signals; with cable lengths exceeding 10 m, we recommend
using R=120 ohms termination resistors between wires (rd) and (wh), (bu and (bk), (gy)
and (gn).
2)
Color code brown (br) for Kab 163 and Kab 164.
Assignment for plug 3:
Supply voltage and voltage output signal.
61
5
72
4
3
Binder 423
device plug
Top view
Plug
pin
Assignment
1
Torque/rotational speed measurement signal (voltage output; 0 V )
or rotational speed measurement signal (0 V)
2
Supply voltage 0 V;
3 Supply voltage 18 V to 30 V DC
4
Torque measurement signal (voltage output; "10 V)
or rotational speed measurement signal ("10 V)
5 Not in use
6 Shunt signal trigger 5 V to 30 V and TEDS for torque
7
Shunt signal 0 V;
Shielding connected to housing ground