User's Manual
Table Of Contents
- T12
- Contents
- Safety instructions
- 1 Markings used
- 2 Scope of supply
- 3 Operation
- 4 Application
- 5 Signal flow
- 6 Structure and mode of operation
- 7 Mechanical installation
- 7.1 Important precautions during installation
- 7.2 Conditions on site
- 7.3 Mounting position
- 7.4 Installing the slotted disc (rotational speed measuring system only)
- 7.5 Installing the rotor
- 7.6 Fitting the protection against contact (option)
- 7.7 Installing the stator
- 7.8 Optical rotational speed/angle of rotation measuring system (option)
- 8 LED status display
- 9 Electrical connection
- 10 Shunt signal
- 11 Load‐carrying capacity
- 12 TEDS
- 13 Maintenance
- 14 Waste disposal and environmental protection
- 15 Specifications
- 16 Dimensions
- 16.1 Rotor 100 N⋅m to 200 N⋅m
- 16.2 Rotor 500 N⋅m to 10 kN⋅m
- 16.3 Stator 100 N⋅m to 200 N⋅m with rot.speed meas. system
- 16.4 Stator 100 N⋅m to 200 N⋅m with rot. speed meas. system
- 16.5 Stator 100 N⋅m to 10 kN⋅m with rot. speed meas. system
- 16.6 Stator 100 N⋅m to 200 N⋅m with prot. against contact
- 16.7 Stator 100 N⋅m to 200 N⋅m with prot. against contact
- 16.8 Stator 500 N⋅m to 1 kN⋅m with prot. against contact
- 16.9 Stator 2 kN⋅m to 10 kN⋅m with prot. against contact
- 16.10 Mounting dimensions
- 17 Supplementary technical information
- 18 Condition at the time of delivery
- 19 Ordering numbers
- 20 Accessories
45
T12
A1979−10.0 en HBM
10 Shunt signal
The T12 torque transducer supplies a shunt signal, at either 50% or 10% of
the nominal (rated) torque, as selected. Activate this function via the T12
Assistant or the shunt signal trigger on plug 1 or plug 3 (see Section 9.3). The
last shunt selected in the T12 Assistant is then triggered.
The internal signal conditioning may cause a delay in triggering of about 5
seconds.
To obtain stable conditions, we recommend activating the shunt signal only
once the transducer has been warming up for 15 minutes.
The framework conditions for reproducibility (e.g. the mounting conditions)
must be established in order to reproduce the measured values in the
manufacturing certificate.
Important
The transducer should not be under load when the shunt signal is being
measured, as the signal is applied additively.
After about 5 minutes, the shunt signal is automatically deactivated.