User's Manual

Table Of Contents
Mechanical installation
T40FM A3276-8.0 HBM: public 27
5.6 Mounting the rotor
Tip
Usually the rotor type plate is no longer visible after in
stallation. This is why we include with the rotor additional
stickers with the important characteristics, which you can
attach to the stator or any other relevant test-bench com
ponents. You can then refer to them whenever there is
anything you wish to know, such as the shunt signal. To
explicitly assign the data, the identification number and
the size are engraved on the rotor flange, where they can
be seen from outside.
Notice
Make sure during installation that you do not damage the
measuring zone marked in Fig. 5.4 by using it to support
tools or knocking tools against it when tightening screws,
for example. This can damage the transducer and pro
duce measurement errors, or even destroy the trans
ducer.
1. Prior to installation, clean the plane faces of the trans
ducer flange and the counter flange.
For safe torque transfer, the faces must be clean and
free from grease. Use a piece of cloth or paper
soaked in solvent. When cleaning, make sure that you
do not damage the transmitter winding.