6 Electrical Connection
6.1 General information
S With extension cables, make sure that there is a
proper connection with minimum contact resistance
and good insulation.
S All cable connectors or swivel nuts must be fully
tightened.
Important
Transducer connection cables from HBM with plugs at
tached are identified in accordance with their intended
purpose (Md or n). When cables are shortened, inserted
into cable ducts or installed in control cabinets, this identi
fication can get lost or become concealed. So the cables
must be marked beforehand, just in case.
6.2 EMC protection
Important
The transducers are EMC-tested in accordance with EC
directives and identified by CE certification. However, you
must connect the shield of the connection cable on the
shielding electronics enclosure in order to achieve EMC
protection for the measuring chain.