User's Manual
Table Of Contents
- T40FM
- T40FM
- English
- 1 Safety information
- 2 Markings used
- 3 Applications
- 4 Structure and Mode of Operation
- 5 Mechanical installation
- 6 Electrical Connection
- 7 Shunt signal
- 8 Functionality testing
- 9 Loading capacity
- 10 Maintenance
- 11 Waste disposal and environmental protection
- 12 Dimensions
- 13 Order numbers, Accessories
- 14 Specifications
- 15 Supplementary Technical Information
- English
- T40FM
- Deutsch
- 1 Sicherheitshinweise
- 2 Verwendete Kennzeichnungen
- 3 Anwendung
- 4 Aufbau und Wirkungsweise
- 5 Mechanischer Einbau
- 6 Elektrischer Anschluss
- 7 Shuntsignal
- 8 Funktionsprüfung
- 9 Belastbarkeit
- 10 Wartung
- 11 Entsorgung und Umweltschutz
- 12 Abmessungen
- 13 Bestellnummern, Zubehör
- 14 Technische Daten
- 15 Ergänzende technische Informationen
- Deutsch
Shunt signal
52 A3276-8.0 HBM: public T40FM
7 Shunt signal
The T40FM torque flange delivers an electrical shunt sig
nal that can be activated from the amplifier for measuring
chains with HBM components. The transducer generates
a shunt signal of about 50% of the nominal (rated) torque;
the precise value is specified on the type plate. After ac
tivation, adjust the amplifier output signal to the shunt
signal supplied by the connected transducer, to adapt the
amplifier to the transducer.
Information
The transducer should not be under load when the shunt
signal is being measured, as the shunt signal is mixed
additively.
Triggering the shunt signal
Applying a separated extra-low voltage of 5 … 30 V to
pins 6 (+) and 7 (
) at plug 1 or 3, triggers the shunt
signal.
The nominal (rated) voltage for triggering the shunt signal
is 5 V (triggering at U > 2.5 V), but when voltages are
less than 0.7 V, the transducer is in measuring mode.
The maximum permissible voltage is 30 V, current con
sumption at nominal (rated) voltage is approx. 2 mA and
at maximum voltage, approx. 18 mA. The voltage for trig
gering the shunt signal is electrically isolated from the
supply and measuring voltage.
Tip
The shunt signal can be triggered by the amplifier or via
the operating software in HBM system solutions.