user manual
CALibrationn:CONFigure:VOLTage?
CALibrationn:CONFigure:VOLTage? configures the channel for voltage mode
calibration and returns a
0 or 1 to indicate the channel’s readiness for calibration.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
CALibration
n keyword 1 to 8 or 1 to 16 1
Returns: an unsigned 16-bit integer (uint16). 1 = ready, 0 = not ready.
Comments • In the keyword, CALibrationn, the valid range of n is:
1 to 8 for E1418A, E1418A Option 002
1 to 16 for E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
• This command is used together with the CALibrationn:VALue:VOLTage? query to
perform voltage mode calibration.
• The calibration is performed either at the CAL output terminals or at an individual
channel. The choice is made using the
DIAGnostic:CALibration:OUTPut[:MODE]
command.
– When the DIAG:CAL:OUTP:MODE CALBus command (also the reset state)
is used, the output relay routes the calibration voltage to the CAL output
terminals (and opens all channel output terminals).
– When the command DIAG:CAL:OUTP:MODE CHANnel is used, the output
relay routes the calibration voltage to the channel terminals (the CAL terminals
are not connected).
The
DIAGnostic:CALibration:OUTPut[:MODE] command must be sent to the
module before the
CALibrationn:CONFigure:VOLTage? query.
• This query will generate an error if the trigger system is waiting for a trigger event
when this command is received.
• This query returns a 0 to indicate that the channel is not ready for calibration,
and a
1 to indicate the channel is ready for calibration.
• This command is affected by the state of the program jumper for this channel on
the terminal module. To use this command:
– the P/J jumper must be set to program selectable or, if set to jumper selectable,
– the V/I jumper must be set to the voltage mode.
The jumpers are described on page 31.
• Refer to Appendix D, “Voltage/Current Output Adjustment”, starting on page 154,
for additional information about this command.
• Related Commands: CALibrationn:VALue:VOLTage?,
DIAGnostic:CALibration:OUTPut[:MODE]
CALibration
72 HP E1418A SCPI Command Reference Chapter 3