user manual

SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude]
SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude] <value | MIN | MAX | DEF>
immediately outputs the current set in value at channel n when the output relay is closed.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
SOURce
n keyword 1 to 8 or 1 to 16 1
<
value>
numeric (float 32) or
discrete
-0.02 Amps to +0.02 Amps or
MIN | MAX | DEF
0
Comments In the keyword, SOURcen, the valid range of n is:
1 to 8 for E1418A, E1418A Option 002
1 to 16 for E1418A Option 001, E1418A Option 003.
Values of n outside these ranges will generate an error.
The <value> parameter is a floating point 32-bit number in the range of -0.02 A
to +0.02 A.
MIN sets -0.02 Amps.
MAX sets +0.02 Amps.
DEF sets 0 Amps.
This command immediately programs the specified current on channel n. The
output state of the channel must be enabled for the current to appear at the output
terminals.
If the channel is in the voltage mode when this command is received, an error is
generated.
Related Commands: SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude]?,
SOURcen:FUNCtion:MODE CURRent, OUTPutn[:STATe]
SOURce
Chapter 3 HP E1418A SCPI Command Reference 85