book.
book.book Page ii Monday, January 31, 2000 10:34 AM © Copyright Hewlett-Packard Company 2000 All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under copyright laws. HP Part No. 86140-90035 Printed in USA February 2000 Hewlett-Packard Company Lightwave Division 1400 Fountaingrove Parkway Santa Rosa, CA 95403-1799, USA (707) 577-1400 Notice. The information contained in this document is subject to change without notice.
book.book Page iii Monday, January 31, 2000 10:34 AM General Safety Considerations General Safety Considerations This product has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
book.book Page iv Monday, January 31, 2000 10:34 AM General Safety Considerations CAUTION Do not use too much liquid in cleaning the optical spectrum analyzer. Water can enter the front-panel keyboard, damaging sensitive electronic components. CAUTION VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection into and out of the product must not be restricted.
book.
book.
book.
book.book Page 2 Monday, January 31, 2000 10:34 AM Setting Up the OSA Setting Up the OSA Setting Up the OSA This chapter shows you how to set up your optical spectrum analyzer, connect power and accessories, and verify general operation. Although the pictures in this section show an HP 86140A or 86142A optical spectrum analyzer, setting up other HP 86140-series optical spectrum analyzers is very similar.
book.book Page 3 Monday, January 31, 2000 10:34 AM Setting Up the OSA Setting Up the OSA Package contents for HP 86140 series optical spectrum analyzers ❒ Inspect the shipping container for damage. ❒ Inspect the instrument. ❒ Verify that you received the options and accessories you ordered. Keep the shipping container and cushioning material until you have inspected the contents of the shipment for completeness and have checked the optical spectrum analyzer mechanically and electrically.
book.book Page 4 Monday, January 31, 2000 10:34 AM Setting Up the OSA Setting Up the OSA You can tilt your instrument upward for easier viewing. On the HP 86140A and 86142A instruments, lift up the front of the instrument, grasp one of the wire bails under the front corner, and pull it down and forward until it latches into place. Repeat for the other wire bail. On HP 86143A and 86145A instruments, pivot the handle to tilt the instrument.
book.book Page 5 Monday, January 31, 2000 10:34 AM Setting Up the OSA Setting Up the OSA Although you can operate all instrument functions using only the front-panel keys and trackball, these accessories make your optical spectrum analyzer even easier to use! Connect any standard PC-compatible mouse (or other pointing device), keyboard, or external VGA-compatible display. CAUTION Do not stack other objects on the keyboard; this will cause self-test failures on power-on.
book.book Page 6 Monday, January 31, 2000 10:34 AM Setting Up the OSA Setting Up the OSA You can connect a PCL-language printer (for example, an HP LaserJet) to the instrument’s rear panel Parallel connector. Use a parallel Centronics printer cable, such as an HP C2950A (2 m) or HP C2951A (3 m).
book.book Page 7 Monday, January 31, 2000 10:34 AM Setting Up the OSA Setting Up the OSA The optical spectrum analyzer automatically adjusts for line input voltages in the range of 100 to 240 VAC. There is no manual selection switch. The line cord provided is matched by HP to the country of origin of the order. Refer to “AC Line-Power Cords” on page 6-4. Line Power Requirements Power 115 VAC: 110 VA MAX. / 60 WATTS MAX. / 1.1 A MAX. 230 VAC: 150 VA MAX. / 70 WATTS MAX. / 0.6 A MAX.
book.book Page 8 Monday, January 31, 2000 10:34 AM Setting Up the OSA Setting Up the OSA • Press the power switch at the lower left-hand corner of the front panel. After a short initialization period, the display will look similar to the picture on this page. The instrument is ready to use.
book.book Page 9 Monday, January 31, 2000 10:34 AM Setting Up the OSA Setting Up the OSA CAUTION Fiber-optic connectors are easily damaged when connected to dirty or damaged cables and accessories. The HP 86140 series’s front-panel INPUT connector is no exception. When you use improper cleaning and handling techniques, you risk expensive instrument repairs, damaged cables, and compromised measurements.
book.
book.
book.book Page 2 Monday, January 31, 2000 10:34 AM A Quick Tour A Quick Tour A Quick Tour HP 86140A-series optical spectrum analyzers display input light spectra from 600 nm to 1700 nm. To change instrument settings use the softkeys. To display different softkeys use these items: • Softkey panel and softkeys • Print, Save/Recall, System, and Application keys • Menu bar The menu bar or keys can be used to access the same functions.
book.book Page 3 Monday, January 31, 2000 10:34 AM A Quick Tour A Quick Tour All displayed wavelength values show values as measured in a vacuum. This default setting can be changed to show values as measured in air. See “To change the default wavelength settings” on page 3-63 for more information. The OPTICAL INPUT connector on standard instrument versions uses single mode fiber. Connecting multi-mode fiber to these connectors results in large reflections and insertion loss.
book.
book.
book.
book.
book.
book.
book.book Page 10 Monday, January 31, 2000 10:34 AM A Quick Tour The Menu Bar The Menu Bar The Menu bar includes the File, Measure, Application, and Options drop-down menus. Each menu selection includes a descriptive label. (Action) Indicates the selection will perform an action such as making a measurement or printing the display. (Panel) Indicates the selection will open a softkey panel.
book.book Page 11 Monday, January 31, 2000 10:34 AM A Quick Tour The Softkey Panels The Softkey Panels The softkey panels can be accessed using either the front-panel keys or the menu bar. This section includes brief descriptions of the following menus. See Chapter 3, “Using the Optical Spectrum Analyzer” for additional information on each of the OSA functions.
book.book Page 12 Monday, January 31, 2000 10:34 AM A Quick Tour The Softkey Panels The Amplitude menus The Amplitude softkeys are accessed by using the front-panel Amplitude key or the Measure menu Amplitude selection on the menu bar.
book.book Page 13 Monday, January 31, 2000 10:34 AM A Quick Tour The Softkey Panels The Applications menus The Applications (Appl’s) softkeys are accessed by using the front-panel Appl’s key or the Applications menu Launch an Installed Application.... selection on the menu bar. For a complete description of the applications, refer to the manual that came with your software.
book.book Page 14 Monday, January 31, 2000 10:34 AM A Quick Tour The Softkey Panels The Bandwidth/Sweep menus The Bandwidth/Sweep softkeys are accessed by using the front-panel Bandwidth/Sweep key or the Measure menu Bandwidth/Sweep selection on the menu bar.
book.book Page 15 Monday, January 31, 2000 10:34 AM A Quick Tour The Softkey Panels The Marker menus The Markers softkeys are accessed using the front-panel Markers key or the Measure menu Markers selection on the menu bar.
book.book Page 16 Monday, January 31, 2000 10:34 AM A Quick Tour The Softkey Panels The Save/Recall menus The Save/Recall softkeys and setup panels are accessed by using the dropdown File menu Save/Recall selection or the front-panel Save/Recall key. Use these functions to save, recall and print the measurement results.
book.book Page 17 Monday, January 31, 2000 10:34 AM A Quick Tour The Softkey Panels The Systems menus The System softkeys are accessed using the front-panel System key or the Options menu System selection on the menu bar.
book.book Page 18 Monday, January 31, 2000 10:34 AM A Quick Tour The Softkey Panels The Traces menus The Traces softkeys are accessed by using the front-panel Traces key or the Measure menu Traces selection on the menu bar.
book.book Page 19 Monday, January 31, 2000 10:34 AM A Quick Tour The Softkey Panels The Wavelength menus The Wavelength softkeys are accessed using the front-panel Wavelength key or the Measure menu Wavelength selection on the menu bar.
book.book Page 20 Monday, January 31, 2000 10:34 AM A Quick Tour Tutorial: Getting Started Tutorial: Getting Started This tutorial will introduce you to the HP 86140 series optical spectrum analyzer front panel controls. You will perform an auto alignment, perform a peak search, use a delta marker and print the display. Instrument setup A source signal must be present at the input of the optical spectrum analyzer. In this tutorial a Fabry-Perot laser was used as the source.
book.book Page 21 Monday, January 31, 2000 10:34 AM A Quick Tour Tutorial: Getting Started 2 Connect a fiber from the source to the input connector of the optical spectrum analyzer. Be sure to follow the good connector practices described in “Cleaning Connections for Accurate Measurements” on page 6-10. 3 Enable the source. 4 Press the front-panel Auto Align key to optimize the detection of the incoming signal. This takes a few moments to complete.
book.book Page 22 Monday, January 31, 2000 10:34 AM A Quick Tour Tutorial: Getting Started Using the delta marker The optical spectrum analyzer has four types of markers; normal markers, bandwidth markers, delta markers and noise markers. The marker currently being displayed is a normal marker. In the next step we will use it as a delta marker. 6 Press the front-panel Marker key. 7 Press the More Marker Functions.... softkey.
book.
book.book Page 24 Monday, January 31, 2000 10:34 AM A Quick Tour Changing the Printer Paper CAUTION Avoid dropping the coin or screwdriver, used to open the printer door, into the printer assembly. CAUTION Always use HP brand paper to ensure quality printing and long printer life. Order paper as HP part number 9270-1370. CAUTION Never use the printer without printer paper as this can damage the printer head.
book.
book.book Page 2 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer Using the OSA Using the OSA In this chapter, there are sections on both the menus and the front-panel keys. Any of the instrument settings can be changed by using either the front-panel keys or the menu bar selections. Many of the menu selections and front-panel keys display a softkey panel. Settings in softkey panels are changed using the softkeys, data-entry keys, mouse, and trackball.
book.book Page 3 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer Front-Panel Keys Front-Panel Keys The optical spectrum analyzer has front-panel keys that perform a function when pressed. For maximum amplitude accuracy, perform the automatic alignment whenever the optical spectrum analyzer has been moved, subjected to large temperature changes, or following warm-up. This function can also be accessed by choosing the Measure menu Auto Align selection on the menu bar.
book.book Page 4 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer Front-Panel Keys Pressing the Auto Meas button on the front panel of the OSA performs an automatic measurement of the largest signal found in a full span sweep and places a marker at the signal peak. Auto measure requires the connection of an external light source. This can be a broadband or narrowband source.
book.book Page 5 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer Front-Panel Keys Places the optical spectrum analyzer under local control. Enables the front panel keys after the instrument has been under control with an external computer via HP-IB. Sets the optical spectrum analyzer to a known state.
book.
book.book Page 7 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Menus The Menus The front-panel menu keys are used to perform calibrations, setup measurements, set markers, save and recall data, set system preferences, and so forth. Pressing a front-panel menu key displays a softkey menu or setup panel.
book.book Page 8 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Amplitude Menus The Amplitude Menus The Amplitude menus are accessed using the front-panel Amplitude key or the Measure menu Amplitude selection on the menu bar. If you wish to change amplitude settings, the following functions and setup panels are available: Function See....
book.book Page 9 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Amplitude Menus Using the Amplitude functions To change the reference level 1 Press the front-panel Amplitude key. 2 Press the Reference Level softkey. 3 Use the knob, step keys or numeric entry pad to enter the desired value. The reference level value is the amplitude level shown at the reference level position.
book.book Page 10 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Amplitude Menus Note If both the left and right vertical scales are displayed, the scale/division of the active trace is changed. To change the display mode between log and linear 1 Press the front-panel Amplitude key. 2 Press the Display Mode softkey to toggle between Log and Linear data display. To change the sensitivity 1 Press the front-panel Amplitude key.
book.book Page 11 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Amplitude Menus Note The settings for sensitivity, video bandwidth and sweep time interact. If the sensitivity is set to manual, the video bandwidth and sweep time may be forced to Auto mode. If the video bandwidth is set to manual, the sensitivity and sweep time may be forced to Auto. If the sweep speed is set to manual and is set too fast, the OVER SWEEP indicator will come on in the display area.
book.book Page 12 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Amplitude Menus Setup panel selections Reference Level Position The value selected for the Reference Level Position determines the position of the reference level on the graticule. Setting this value to zero divisions places the reference level on the very bottom of the grid. Setting the reference level to 10 divisions places the reference level at the top of the grid.
book.book Page 13 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Amplitude Menus Note The optical spectrum analyzer performs a more complete zeroing when the instrument is first turned on. You can run this more complete routine anytime by pressing the Zero Now softkey. Refer to “To zero the instrument” on page 3-50 for more information. Auto Chop Mode Enables the spectrum analyzer chop mode. Chop mode increases dynamic range for long sweep times by subtracting ambient light.
book.book Page 14 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Bandwidth/Sweep Menus The Bandwidth/Sweep Menus The Bandwidth/Sweep softkeys are accessed by using the front-panel Bandwidth/Sweep key or the Measure menu Bandwidth/Sweep selection on the menu bar. If you wish to change bandwidth or sweep settings, the following functions and setup panels are available: Function See....
book.book Page 15 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Bandwidth/Sweep Menus Using the Bandwidth/Sweep functions To change the resolution bandwidth The ability to display two closely spaced signals as two distinct responses is determined by resolution bandwidth. 1 Press the front-panel Bandwidth/Sweep key. 2 Press the Res BW softkey. 3 Toggle between selecting the resolution bandwidth automatically or manually.
book.book Page 16 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Bandwidth/Sweep Menus The following functions affect video bandwidth: • Changing the sensitivity value. • Changing the reference level. • Turning autoranging ON or OFF . The range of video bandwidth available in Auto mode is much greater than can be set manually from the front panel. A lower video bandwidth value requires a longer sweep time.
book.book Page 17 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Bandwidth/Sweep Menus Manually setting the sweep time turns coupling off. Note If the sweep time is set too fast, an OVER SWEEP message appears indicating the display is no longer calibrated and that trace data may not meet specifications. Increase the sweep time until the OVER SWEEP message disappears. If the sweep time is set too slow, measurement times may be excessively long.
book.book Page 18 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Bandwidth/Sweep Menus To select the internal trigger 1 Press the front-panel Bandwidth/Sweep key. 2 Press the More BW/Sweep Functions.... softkey. 3 Press the Trigger Mode.... softkey. 4 Press the Internal softkey. Internal triggering synchronizes the start of the sweep to an internally generated trigger signal. Internal triggering ensures continuously triggered sweeps with the shortest delay between sweeps.
book.book Page 19 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Bandwidth/Sweep Menus To select an external trigger 1 Press the front-panel Bandwidth/Sweep key. 2 Press the More BW/Sweep Functions.... softkey. 3 Press the Trigger Mode.... softkey. 4 Press the External softkey. External triggering synchronizes the sweep with a signal connected to the rear-panel EXT TRIG IN connector.
book.book Page 20 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Bandwidth/Sweep Menus To set the trigger delay 1 Press the front-panel Bandwidth/Sweep key. 2 Press the More BW/Sweep Functions.... softkey. 3 Press the Trigger Delay softkey. 4 Use the knob, step keys or numeric entry pad to enter the desired value. 5 Use the Trigger Delay softkey to set the delay between the external trigger and the triggering of the analog-to-digital converter.
book.book Page 21 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus The Markers Menus The Markers softkeys are accessed using the front-panel Markers key or the Measure menu Markers selection on the menu bar. If you wish to change marker settings, the following functions and setup panels are available: Function See.... Active Marker.... “To select the active marker” on page 3-23 Marker Trace....
book.book Page 22 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus Function See.... Wavelength Marker 2 “To set the wavelength line markers” on page 3-28 All Line Markers Off “To turn off all line makers and line marker functions” on page 3-29 Advanced Line Marker Functions....
book.book Page 23 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus Using the Marker functions To select the active marker 1 Press the front-panel Markers key. 2 Press the Active Marker.... softkey. 3 Select the appropriate marker to activate, deactivate, or turn all markers off. Selecting a marker always places that marker on the currently selected Active trace at the center wavelength.
book.book Page 24 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus To set markers to measure bandwidth 1 Press the front-panel Markers key. 2 Select an active marker. 3 Place the marker on the peak of the signal to be measured by pressing the PEAK SEARCH softkey. 4 Press the Marker BW.... softkey. 5 Select a pre-defined bandwidth or use the knob, step keys, or numeric entry pad to enter the active marker bandwidth amplitude for the active marker.
book.book Page 25 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus To change the reference level with a marker 1 Press the front-panel Markers key. 2 Place a marker on the trace at the level you want moved to the reference level position. 3 Press the Marker to REF LEVEL softkey. The reference level is set to the amplitude of the active marker. To perform a peak search 1 Press the front-panel Markers key. 2 Select an active marker. 3 Press the PEAK SEARCH softkey.
book.book Page 26 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus Next Peak Right Places the marker on the next peak located at a longer wavelength than the current marker wavelength position. This next peak must meet the peak excursion and threshold criteria. If the specified marker is OFF, it will be turned ON, placed at the center wavelength, and the search to the right will begin from that point. Active Marker.... Turns a particular marker ON or OFF .
book.book Page 27 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus Next Pit Right Places the marker on the next pit located at a longer wavelength than the current marker wavelength position. This next pit must meet the pit excursion and threshold criteria. If the specified marker is OFF , it will be turned ON , placed at the center wavelength, and the search to the right will begin from that point. Active Marker.... Turns a particular marker ON or OFF .
book.book Page 28 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus To set the delta marker function 1 Place the marker to be used as a reference on the desired trace and at the desired wavelength. 2 Press the front-panel Markers key. 3 Press the More Marker Functions.... softkey. 4 Press the Delta Marker softkey. To fix the position of the reference marker and activate the delta marker, toggle the delta marker ON.
book.book Page 29 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus To turn off all line makers and line marker functions 1 Press the front-panel Markers key. 2 Press the More Marker Functions.... softkey. 3 Press the Line Marker Menu.... softkey. 4 Press the Line Markers Off softkey. To sweep between line markers 1 Press the front-panel Markers key. 2 Press the More Marker Functions.... softkey. 3 Press the Line Marker Menu.... softkey.
book.book Page 30 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus To integrate between line markers 1 Press the front-panel Markers key. 2 Press the More Marker Functions.... softkey. 3 Press the Line Marker Menu.... softkey. 4 Select and set Wavelength Marker 1 and 2 to the desired integration limits. 5 Press the Advanced Line Mkr Functions.... softkey. 6 Press the Integrate Limit softkey to toggle this function ON or OFF .
book.book Page 31 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus The Marker Setup panel Setup panel selections Normal Marker Units Sets the X-axis readout for frequency or wavelength when the instrument is in a non-zero span. This setting controls only the normal marker X-axis and the delta reference readout. This setting controls all four normal markers. The bandwidth and delta offset markers have their own settings.
book.book Page 32 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus Bandwidth Marker Interpolation Turns the bandwidth marker interpolation ON or OFF . When interpolation is ON , the bandwidth markers will be placed at the exact number of dB (NDB) from the normal marker if within the trace range. The position of the marker will be linearly interpolated between two trace data points. The default state is ON .
book.book Page 33 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Markers Menus Use Marker Search Threshold This limits the marker search function to data points above the selected threshold level. Marker Search Threshold Value Selects the amplitude threshold value used for limiting the marker search function. When Marker Search Threshold is ON, a dotted line is shown on the display at the threshold level.
book.book Page 34 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Save/Recall Menus The Save/Recall Menus The Save/Recall softkeys are accessed by using the front-panel Save/Recall key or the File menu Save/Recall selection on the menu bar. If you wish to change save or recall settings, the following functions are available: Function See.... Save Menu.... “To save measurement and trace data” on page 3-35 Recall Menu....
book.book Page 35 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Save/Recall Menus Using the Save/Recall functions To save measurement and trace data 1 Press the front-panel Save/Recall key. 2 Press the Save Menu.... softkey. 3 The Save Setup panel opens. Refer to “To Fill In a Setup Panel” on page 3-65 for information on changing and selecting items in the setup panel. The Save Setup panel Setup panel selections Save Saves the measurement data in a binary format (.dat file).
book.book Page 36 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Save/Recall Menus Save to Selects saving data to a floppy disk or to internal memory. File Name Selects manual or automatic mode for choosing a file name. 4 When you are satisfied with your selections, press the Choose File to Save softkey. The Filename Menu setup panel opens.
book.book Page 37 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Save/Recall Menus To recall measurement and trace data 1 Press the front-panel Save/Recall key. 2 Press the Recall Menu.... softkey. 3 The Recall Setup panel opens. The Recall Menu setup panel Setup panel selections Recall Selects whether a measurement or trace will be recalled. Recall From Selects whether to recall from a floppy disk or from internal memory.
book.book Page 38 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Save/Recall Menus The Catalog setup panel 6 Use the arrow keys to highlight the next five previous file softkeys, or click on the desired filename using a trackball or mouse. Press Enter to recall or delete the selected file. To delete a file from internal memory or a floppy disk 1 Press the front-panel Save/Recall key. 2 Press the Delete Menu.... softkey.
book.book Page 39 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Save/Recall Menus To format a floppy disk 1 Press the front-panel Save/Recall key. 2 Press the Format Floppy Disk.... softkey. 3 Press the Format softkey to format the 3.5 floppy disk. Press the Cancel softkey to stop the format action and return to the previous screen. To backup or restore the internal memory 1 Press the front-panel Save/Recall key. 2 Press the Backup/Restore Menu.... softkey.
book.book Page 40 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus The System Menus The System softkeys are accessed using the front-panel System key or the Options menu System selection on the menu bar. If you wish to change system settings, the front-panel System key provides access to the following functions and setup panels: Function See.... Help.... Show Critical Errors....
book.book Page 41 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus Function See.... More System Functions.... OSA State.... “To display the OSA State information panel” on page 3-47 Active Panel “To display the Active Function Area” on page 3-48 Set Time/Date.... “To change the time and date” on page 3-48 Service Menu....
book.book Page 42 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus Using the System functions To display notices, bandwidth errors, critical errors and warnings Note The Show Notices...., Show BW Errors...., Show Critical Errors...., and Show Warnings.... functions are only available if a notice, bandwidth error, critical error, or warning has been generated by the system.
book.book Page 43 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus The Title Setup panel Refer to “To Fill In a Setup Panel” on page 3-65 for information on changing and selecting items in the setup panel. To change the display setup 1 Press the front-panel System key. 2 Press the Display Setup.... softkey. The Display Setup panel appears.
book.book Page 44 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus Active Function Area Assist When this function is ON, the function set in the Active Function Area is automatically set to the first, or top, softkey function. For example, when you press the front-panel Bandwidth/Sweep key, the Active Function Area Assist function will set the Active Function Area to resolution bandwidth. This means you can immediately enter the desired resolution bandwidth.
book.book Page 45 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus Setup panel selections Factory Power Cal Date Shows the date of the last factory-performed power calibration. User Power Cal Date Shows the date of the last successful user-performed power calibration. Set Calibration Power This is the exact amplitude of the calibration source that will be used for the next power calibration.
book.book Page 46 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus The Wavelength Calibration setup panel Setup panel selections Factory Wavelength Cal Date Shows the date of the last factory-performed wavelength calibration. User Wavelength Cal Date Shows the date of the last successful user-performed wavelength calibration. Signal Source Select either an external single wavelength signal source or the internal calibrator as the wavelength calibration source.
book.book Page 47 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus 5 Press the Perform Calibration.... softkey. 6 Follow the on-screen instructions and then press Execute Calibration. To move the active function area 1 Press the front-panel System key. 2 Press the Move Active Area softkey. Each press of the softkey moves the active function area to one of eight on-screen locations. To display the OSA State information panel 1 Press the front-panel System key.
book.book Page 48 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus To display the Active Function Area 1 Press the front-panel System key. 2 Press the More System Functions.... softkey. 3 Press the Active Panel softkey to toggle the display of the Active Function Area ON or OFF . To change the time and date NOTE Changing the time, date, or time zone causes the instrument to reboot. 1 Press the front-panel System key. 2 Press the More System Functions.... softkey.
book.book Page 49 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus The IP function is the same as the front-panel Preset key. For a complete list of the Preset conditions, see page 3-5. When Last is selected the instrument will power-on in the same state it was in when last powered-off. To perform a factory preset (IP) 1 Press the front-panel System key. 2 Press the More System Functions.... softkey. 3 Press the Service Menu.... softkey.
book.book Page 50 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus Function Factory Preset Value Miscellaneous functions Power-on in last state HP-IB address Time display HP logo display Active function area assist Active function area location Printer interface Off 23 On On On Upper right Internal To upgrade the firmware To upgrade the firmware you must have the firmware upgrade disk set from HP. 1 Press the front-panel System key.
book.book Page 51 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus Note The optical spectrum analyzer compensates for temperature-related current drift between each sweep. Although this “zeroing” increases amplitude accuracy, it increases the time between sweeps. If amplitude accuracy is not critical to your measurement, you can turn off zeroing. Refer to “Auto Zero” on page 3-12 for additional information.
book.book Page 52 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus 5 Press the Wavelength Limit softkey. Toggle ON to limit the span to the specified range of 600 to 1700 nm. When this function is OFF, the start wavelength of the OSA can be tuned down to 350 nm, and the stop wavelength can be tuned up to 2000 nm. Note The performance of the OSA is not specified and the amplitude is not calibrated outside the 600–1700 nm range.
book.book Page 53 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus To display the OSA Extended State information panel 1 Press the front-panel System key. 2 Press the More System Functions.... softkey. 3 Press the Service Menu.... softkey. 4 Press the Adv Service Functions.... softkey. 5 Press the More Adv Service Menu.... softkey. 6 Press the OSA Extended State.... softkey. The OSA Extended State Information panel is displayed and can be printed.
book.book Page 54 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus 4 Use the arrow keys or the numeric entry pad to set the wavelength span manually or automatically. Refer to “To Fill In a Setup Panel” on page 3-65 for information on changing and selecting items in the setup panel. Setup panel selections Span This function selects the wavelength span for viewing the signal located by the auto measure function.
book.book Page 55 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The System Menus Note None of the Automeasure Setup parameters are affected by the front-panel Preset key. They are not saved as part of the measurement setup. To set the HP-IB address 1 Press the front-panel System key. 2 Press the More System Functions.... softkey. 3 Press the Remote Setup.... softkey. The Remote Setup panel appears.
book.book Page 56 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Traces Menus The Traces Menus The Traces softkeys are accessed by using the front-panel Traces key or the Measure menu Traces selection on the menu bar. If you wish to change trace settings, the front-panel Traces key provides access to the following functions and setup panels: Function See.... Active Trace....
book.book Page 57 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Traces Menus Using the Traces functions To set the active trace 1 Press the front-panel Traces key. 2 Press the Active Trace.... softkey. 3 Use the arrow keys or knob to select the active trace. or 1 Press the front-panel Markers key. 2 Press the Active Trace.... softkey. 3 Use the arrow keys or knob to select the active trace. To put a marker on a trace, or to make changes to a trace, it must be the active trace.
book.book Page 58 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Traces Menus To set the Hold.... function 1 Press the front-panel Traces key. 2 Press the Hold .... softkey to select None, Min or Max. None turns the Hold function off. The maximum hold function compares the current amplitude value of each point on a trace in the current sweep to the corresponding point detected during the previous sweep, then displays the maximum value.
book.book Page 59 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Traces Menus To set the trace math for trace F 1 Press the front-panel Traces key. 2 Press the Trace Math.... softkey. 3 Press the Default Math Trace F.... softkey. The result is placed in trace F. 4 Select the appropriate softkey to define the math expression. F=CLOG –D Defines the math expression to be used and turns the math operation ON. The math operation is performed in linear units.
book.book Page 60 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Traces Menus To set up averaging 1 Press the front-panel Traces key. 2 Press the Averaging.... softkey. 3 Toggle averaging ON or OFF . If averaging is on, softkeys appear to select the number of measurements to be averaged. Use the keys to select the number of measurements to be averaged.
book.book Page 61 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Wavelength Menus The Wavelength Menus The Wavelength softkeys are accessed using the front-panel Wavelength key or the Measure menu Wavelength selection on the menu bar. If you wish to change the wavelength settings, the front-panel Wavelength key provides access to the following functions and setup panels: Function See....
book.book Page 62 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Wavelength Menus Using the Wavelength functions To set the center wavelength 1 Press the front-panel Wavelength key. 2 Press the Center WL softkey. 3 Use the knob, step keys or numeric entry pad to enter the desired value. To set the wavelength span 1 Press the front-panel Wavelength key. 2 Press the Span softkey. 3 Use the knob, step keys or numeric entry pad to enter the desired value.
book.book Page 63 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Wavelength Menus To set the peak to the center of the display 1 Press the front-panel Wavelength key. 2 Press the Peak to CENTER softkey to move the peak to the center of the display. To change the default wavelength settings 1 Press the front-panel Wavelength key. 2 Press the Wavelength Setup.... softkey. The Wavelength Setup panel opens.
book.book Page 64 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer The Wavelength Menus Wavelength Step Size Specifies the center wavelength step size. Wavelengths Referenced In Specifies if the displayed wavelengths show values as measured in air or in vacuum. Pressing the front-panel Preset key sets this value to Vacuum. See page 3-5 for more information on the Preset states.
book.book Page 65 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer To Fill In a Setup Panel To Fill In a Setup Panel Setup panels allow you to adjust setup conditions which are not frequently changed. Using the softkeys The arrow softkeys Allow the user to navigate from field to field in the dialog box. The highlighted parameter can be changed. The Select softkey Selects or deselects the highlighted parameter. The Defaults softkey Resets the parameters to their default condition.
book.book Page 66 Monday, January 31, 2000 10:34 AM Using the Optical Spectrum Analyzer To Fill In a Setup Panel To use the navigation softkeys 1 Use the arrow softkeys to highlight the settings on the setup panel. 2 Use the Select softkey to toggle the selection boxes on and off. 3 Enter values in the numeric fields using the front-panel knob or numeric entry pad. 4 To return the setup values to the instrument’s preset settings, press the Defaults softkey.
book.
book.book Page 2 Monday, January 31, 2000 10:34 AM Remote Operation Remote Operation Remote Operation This chapter provides information on remote programming of the HP 86140series optical spectrum analyzers. The easiest way to program the instrument is by using the VXIplug&play universal instrument drivers. The VXIplug&play universal instrument drivers allow you to develop programs using the following applications: HP VEE, LabVIEW 1, LabWindows/CVI, C, C++, and Microsoft®2 Visual Basic3.
book.book Page 3 Monday, January 31, 2000 10:34 AM Remote Operation Remote Operation Table 4-1. Front-Panel Features Indicates that the instrument is operating under HP-IB control. Indicates that the instrument has requested service from the computer. Refer to “Monitoring the Instrument” on page 4-11. Pressing this button activates the front-panel keys after a computer has had control of the instrument.
book.book Page 4 Monday, January 31, 2000 10:34 AM Remote Operation Getting Started Getting Started One of the easiest ways to learn how to write programs to control the instrument is to look at simple examples. In “Example Programs” on page 4-16, you’ll find several useful example programs. Although they are written using the HP BASIC language, you can easily convert them to the language that you are using. The HP 86140 series’ HP-IB address is configured at the factory to a value of 23.
book.book Page 5 Monday, January 31, 2000 10:34 AM Remote Operation Getting Started Remote command buffering The OSA accepts serial data via HP-IB, testing each byte. Once a complete command is received and interpreted, the HP-IB handshake is held until the command operation is completed. Once completed, the next command byte is read by the analyzer.
book.book Page 6 Monday, January 31, 2000 10:34 AM Remote Operation Getting Started Syntax notation conventions The following information applies to the common and instrument-specific commands listed in this chapter. All measurement values and parameters are sent and received as ASCII strings with the exception of the following commands. These commands send and receive floating point binary data in IEEE 488.
book.book Page 7 Monday, January 31, 2000 10:34 AM Remote Operation Getting Started SCPI command are grouped in subsytems In accordance with IEEE 488.2, the instrument’s commands are grouped into “subsystems.” Commands in each subsystem perform similar tasks. The first page of this chapter lists where each subsystem is documented. Sending a command It’s easy to send a command to the instrument.
book.book Page 8 Monday, January 31, 2000 10:34 AM Remote Operation Getting Started Combine commands from different subsystems You can send commands and program queries from different subsystems on the same line; simply precede the new subsystem by a semicolon followed by a colon. In the following example, the colon and semicolon pair before CALC allows you to send a command from another subsystem.
book.book Page 9 Monday, January 31, 2000 10:34 AM Remote Operation Getting Started Numbers All numbers are expected to be strings of ASCII characters. Thus, when sending the number 9, you would send a byte representing the ASCII code for the character “9” (which is 57). A three-digit number like 102 would take up three bytes (ASCII codes 49, 48, and 50). This is taken care of automatically when you include the entire instruction in a string. Several representations of a number are possible.
book.book Page 10 Monday, January 31, 2000 10:34 AM Remote Operation Getting Started Querying data Data is requested from the instrument using a query. Queries can be used to find out how the instrument is currently configured. They are also used to obtain results of measurements made by the instrument, with the query actually activating the measurement. String responses are returned as uppercase letters. Queries usually take the form of a command followed by a question mark (?).
book.book Page 11 Monday, January 31, 2000 10:34 AM Remote Operation Monitoring the Instrument Monitoring the Instrument Your programs can monitor the HP 86140 series for its operating status, including querying execution or command errors and determining whether or not measurements have been completed. Several status registers and queues are provided to accomplish these tasks as shown in Figure 4-1 on page 4-12.
book.book Page 12 Monday, January 31, 2000 10:34 AM Remote Operation Monitoring the Instrument Figure 4-1.
book.book Page 13 Monday, January 31, 2000 10:34 AM Remote Operation Monitoring the Instrument The STATus:PRESet command clears all event registers and sets all bits in the event enable registers. Use the *CLS common command to clear all event registers and all queues except the output queue. If *CLS is sent immediately following a program message terminator, the output queue is also cleared. In addition, the request for the *OPC bit is also cleared.
book.book Page 14 Monday, January 31, 2000 10:34 AM Remote Operation Monitoring the Instrument Operation Status Structure Contains bits that report on the instrument’s normal operation. Additional transition filters give you the ability to select the logic transitions which set the event register bits. For example, you can define the Measuring bit of the Operation Status Structure to be set when a status line transitions from false to true. This is a positive transition.
book.book Page 15 Monday, January 31, 2000 10:34 AM Remote Operation Monitoring the Instrument Output Queue The output queue stores the instrument responses that are generated by certain commands and queries that you send to the instrument. The output queue generates the Message Available Summary bit when the output queue contains one or more bytes. This summary bit sets the MAV bit (bit 4) in the Status Byte. The method used to read the output queue depends upon the programming language and environment.
book.book Page 16 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs Example Programs These programs are provided to give you examples of using HP 86140 series remote programming commands in typical applications. They are not meant to teach general programming techniques or provide ready-to-use solutions. They should allow you to see how measurements are performed and how to return data to the computer. The programs are written in HP BASIC for Windows.
book.book Page 17 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs Example 1. Initialization and a simple measurement Description This program provides the basic building block for beginning development of a measurement routine. The *RST common command resets the instrument to predetermined settings to provide a common starting point. The automeasure function locates the largest signal in the spectrum and optimizes the display of the signal.
book.book Page 18 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs Example 2. Locating the largest signal Description This program finds the largest signal, zooms to a narrow span, and then uses markers to return signal wavelength and amplitude to the computer.
book.book Page 19 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs Example 3. Bandwidth Description The 20 dB marker BW function is used to determine the bandwidth of the signal. The program assumes a narrowband signal as an input.
book.book Page 20 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs OUTPUT 723;”calc1:mark1:func:bwid:read wav” OUTPUT 723;”calc1:mark1:func:bwid:stat on” ! OUTPUT 723;”calc1:mark1:func:bwid:res?” ENTER 723;Rbw IF Rbw<9.E+37 THEN Cnt PRINT “ BW not found” STOP Cnt: ! OUTPUT 723;”calc1:mark1:func:bwid:x:left?” ENTER 723;Markleft OUTPUT 723;”calc1:mark1:func:bwid:x:righ?” ENTER 723;Markright ! ! Convert to standard measurement units (nm) Rbw=Rbw*1.E+9 Markleft=Markleft*1.
book.book Page 21 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs Example 4. Maximum and minimum amplitude values Description The marker delta function is used to find the maximum and minimum (peak and pit) values of the signal.
book.book Page 22 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs PRINT “ Marker values” PRINT Markwl=Markwl*1.E+9 PRINT Markwl;”nm”,Markamp;”dBm” ! Markdeltx=Markdeltx*1.
book.book Page 23 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs Example 5. Maximum and minimum values over time Description This program locates the largest signal using automeasure, and adjusts the center wavelength, span, and sensitivity settings. Trace B is then viewed and updated and set to maximum hold. Trace C is then viewed, updated, and set to minimum hold. Signal variations with time can now be monitored.
book.book Page 24 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs Example 6. Returning trace data Description This program locates the largest signal and then zooms to a narrow span. The trace length is changed to 101 points and the entire trace data is read in and printed on the display.
book.book Page 25 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs ENTER 723;Startw OUTPUT 723;”sens:wav:stop?” ! Read stop wavelength ENTER 723;Stopw OUTPUT 723;”sens:swe:poin?” ! Read trace length ENTER 723;Tlength Bucket=(Stopw-Startw)/(Tlength-1) ! Calculate bucket length PRINT “Data Point Size”,Bucket;”nm” ! PRINT “Point”,” Wavelength”,”Amplitude” ! ! The following lines calculate the wavelength value of each point ! of the trace.
book.book Page 26 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs Example 7. Trace normalization Description This program demonstrates trace normalization. Normalization is used to observe changes to a displayed response. For example, run the program and then bend the fiber to observe the change in signal level across the spectrum. Trace C displays the difference between trace A and trace B.
book.
book.book Page 28 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs Example 8. Total power measurement Description This program demonstrates the total power function. The ASE broadband noise power of an EDFA source is measured. Two sweeps are taken, one of the entire trace and then another of just the noise hump. The total power of the two different traces are displayed.
book.book Page 29 Monday, January 31, 2000 10:34 AM Remote Operation Example Programs Example 9. Monitoring the status registers Description This program presets the analyzer and then selects the largest signal using automeasure. This program demonstrates the use of status registers to detect programming errors. A serial poll is performed to read the instrument status byte. The same status byte is read with *STB?. The internal error register is also read and displayed.
book.
book.book Page 31 Monday, January 31, 2000 10:34 AM Remote Operation Front Panel Functions to Remote Commands Front Panel Functions to Remote Commands This is a table of the front-panel functions of the HP 86140 series and the corresponding remote commands. Table 4-7.
book.book Page 32 Monday, January 31, 2000 10:34 AM Remote Operation Front Panel Functions to Remote Commands Table 4-7.
book.book Page 33 Monday, January 31, 2000 10:34 AM Remote Operation Front Panel Functions to Remote Commands Table 4-7.
book.book Page 34 Monday, January 31, 2000 10:34 AM Remote Operation Front Panel Functions to Remote Commands Table 4-7.
book.book Page 35 Monday, January 31, 2000 10:34 AM Remote Operation Front Panel Functions to Remote Commands Table 4-7.
book.book Page 36 Monday, January 31, 2000 10:34 AM Remote Operation Front Panel Functions to Remote Commands Table 4-7.
book.
book.
book.
book.
book.
book.book Page 42 Monday, January 31, 2000 10:34 AM Remote Operation Common Commands Common Commands *CLS Clears all the event status registers summarized in the status byte register. This command resets the status data structure. It does this by emptying the error queue and clearing all bits in all of the event registers. *ESE Sets the bits in the Standard Event Enable register.
book.book Page 43 Monday, January 31, 2000 10:34 AM Remote Operation Common Commands *OPC Sets bit 0 in the Standard Event Status register when all pending operations have finished. *OPC? Returns a 1 when all operations have finished. *OPT? Returns a comma-separated list of strings that identifies the optical spectrum analyzer's option configuration. A 0 indicates no options are present.
book.book Page 44 Monday, January 31, 2000 10:34 AM Remote Operation Common Commands *STB? Returns the current value of the instrument’s Status Byte. This will not change the Status Byte register. The response value is an integer, to be interpreted as a binary number, representing the bit values of the register. Performing a serial poll on the instrument also reads the Status Byte register, except that bit 6 indicates whether there is a service request that has not been serviced.
book.book Page 45 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate Subsystem Commands The CALCulate subsystem performs post-acquisition data processing. The CALCulate subsystem operates on data acquired by a SENSe function. For more information, refer to page 4-1 of the 1997 SCPI Command Reference. Note CALC: is interpreted as CALC1:. CALC1 controls TRA, CALC2 controls TRB, CALC3 controls TRC, CALC4 controls TRD, CALC5 controls TRE, and CALC6 controls TRF.
book.book Page 46 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate[1|2|3|4|5|6]:AVERage[:STATe] OFF|ON|0|1 CALCulate[1|2|3|4|5|6]:AVERage[:STATe]? Turns trace averaging ON and OFF . If a math expression with the corresponding CALCulate subopcode is OFF , the SENSe:DATA is averaged. If the math expression is ON, the result of the math expression is averaged. The CALCulate:AVERage, CALCulate:MAXimum, and CALCulate:MINimum states are mutually exclusive.
book.book Page 47 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands values closest to and more negative than the NDB value. For positive NDB values, the bandwidth markers will be at values closest to and more positive than the NDB values. This will typically result in a wider bandwidth measurement. This is a global setting and controls the interpolation state for all four bandwidth markers.
book.book Page 48 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer[1|2|3|4]:FUNCtion:BWIDth|BANDwidth[:STATe] OFF|ON|0|1 CALCulate:MARKer[1|2|3|4]:FUNCtion:BWIDth|BANDwidth[:STATe]? Turns the bandwidth marker function ON or OFF for a particular marker. Only one marker function is allowed to be on at a time for each individual marker. Turning on the bandwidth function for a marker will turn off any other marker function and turn on the Bandwidth function.
book.book Page 49 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer[1|2|3|4]:FUNCtion:BWIDth|BANDwidth:X:RIGHt? Returns the absolute X-axis value of the right bandwidth marker. The units returned are determined by the CALCulate:MARKer:FUNCtion:BWIDth|BANDwidth:X:READout state. For READout of FREQuency, the X value is returned in Hertz. For READout of WAVelength, the X value is returned in meters. For READout of WAVelength, the X value is returned in meters.
book.book Page 50 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer[1|2|3|4]:FUNCtion:DELTa:X:OFFSet:FREQuency Allows the user to set the marker offset in frequency units. The marker X-axis value corresponds to the reference X value + the offset value. The default units of the parameter for this command are Hertz. This query generates a “Settings conflict” error if the delta function is OFF for the specified marker.
book.book Page 51 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands Trying to set the readout to TIME when in a non-zero span generates a “Settings conflict” error. Trying to set the readout to FREQuency or WAVelength when in a zero span also generates a “Settings conflict” error. When the instrument is set to zero span, the readout will automatically change to TIME. If the delta marker is OFF a “Settings conflict” error is generated.
book.book Page 52 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer[1|2|3|4]:FUNCtion:NOISe:RESult? This query returns the noise marker value normalized to 1 or 0.1 nm. The normalization bandwidth is controlled by the CALCulate:MARKer:FUNCtion:NOISe:BWIDth command. This query generates a “Settings conflict” error if the noise function is OFF for the specified marker.
book.book Page 53 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer[1|2|3|4]:MAXimum:LEFT Places the marker on the next peak located at a shorter wavelength than the current marker wavelength position. This next peak must meet the peak excursion and threshold criteria. If the specified marker is OFF, it will be turned ON, placed at the center wavelength, and the search to the left will begin from that point.
book.book Page 54 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer[1|2|3|4]:MINimum:NEXT Places the marker on the next lowest pit from the current marker amplitude. This next lowest pit must meet the pit excursion and threshold criteria. If the specified marker is OFF , it will be turned ON, placed at the center wavelength, and the search for the next minimum will begin from that point.
book.book Page 55 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer:SRANge:LOWer:FREQuency Sets the lower limit for the marker search range. Setting this value when CALCulate:MARKer:SRANge:STATe is OFF will automatically turn CALCulate:MARKer:SRANge:STATe ON. The range used for the marker search range is the same range used for the total power calculation, the trace mean range, and the wavelength sweep range.
book.book Page 56 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer:SRANge[:STATe] OFF|ON|0|1 CALCulate:MARKer:SRANge[:STATe]? Turns the search range ON or OFF for all the markers. When the search range is ON , all the marker maximum/minimum searches will be within the upper and lower wavelength range.
book.book Page 57 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer:SRANge:UPPer:TIME Sets the upper limit for the marker search range. Setting this value when CALCulate:MARKer:SRANge:STATe is OFF will automatically turn CALCulate:MARKer:SRANge:STATe ON. The range used for the marker search range is the same range used for the total power calculation, the trace mean range, and the wavelength sweep range.
book.book Page 58 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer[1|2|3|4]:TRACe TRA|TRB|TRC|TRD|TRE|TRF CALCulate:MARKer[1|2|3|4]:TRACe? Places the marker on a particular trace. CALCulate:MARKer[1|2|3|4]:X? Returns the X-axis value of the normal marker. When the delta function is ON , the absolute X-axis value of the delta marker is returned. When the bandwidth function is ON, the X-axis value of the center marker is returned.
book.book Page 59 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate:MARKer[1|2|3|4]:X:TIME Sets the X-axis value of the normal marker when the instrument is in zero span. When the delta function is ON, the absolute X-axis value of the delta marker is controlled. When the bandwidth function is ON, the X-axis value of the center marker is controlled. The default units of the parameter is seconds.
book.book Page 60 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate[1|2|3|4|5|6]:MATH[:EXPRession][:DEFine] () CALCulate[1|2|3|4|5|6]:MATH[:EXPRession][:DEFine]? ::= ::= [] +|–|*|/ Note The parentheses in the above expression are required for proper operation of this command. Defines a math expression to be used when the math operations are turned ON .
book.book Page 61 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate[1|2|3|4|5|6]:MATH:STATe OFF|ON|0|1 CALCulate[1|2|3|4|5|6]:MATH:STATe Determines whether or not math processing is done. CALCulate[1|2|3|4|5|6]:MAXimum:CLEar Clears the current maximum hold values for the trace and allows a new maximum hold to occur. The trace will be initialized to a very negative dBm value (–300 dBm).
book.book Page 62 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate[1|2|3|4|5|6]:MEAN[:DATA]? Returns the arithmetic mean of the trace associated with the subopcode of the CALCulate node. The points of the trace are summed in linear units and the sum is divided by the number of points. When the CALCulate:MEAN:RANge is ON , the mean is calculated over the upper and lower X-axis range limits.
book.book Page 63 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate[1|2|3|4|5|6]:MEAN:RANGe:LOWer:TIME [NS|US|MS|S] Sets the lower X-axis limit for the trace mean range calculation. Setting this value when CALCulate:MEAN:RANGe:STATe is OFF will automatically turn CALCulate:MEAN:RANGe:STATe ON . The range used for the trace mean range is the same range used for the total power calculation, the marker search range, and the wavelength sweep range.
book.book Page 64 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate[1|2|3|4|5|6]:MEAN:RANGe:UPPer? This query returns the upper X-axis limit for the trace mean range calculation. The range used for the trace mean range is the same range used for the total power calculation, the marker search range, and the wavelength sweep range. The returned value is in meters, unless the span is set to zero, in which case the returned value is in seconds.
book.book Page 65 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate[1|2|3|4|5|6]:MEAN:STATe OFF|ON|0|1 CALCulate[1|2|3|4|5|6]:MEAN:STATe? Turns the mean power calculation for a trace ON or OFF . Only one mean power calculation can be turned on at a time. For example, if a mean power calculation is being performed on trace A, turning a mean power calculation for trace B ON will turn the calculation for trace A OFF .
book.book Page 66 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate[1|2|3|4|5|6]:THReshold:STATe ON|OFF|1|0 CALCulate[1|2|3|4|5|6]:THREshold:STATe? Turns on the marker search threshold function. When this threshold function is ON, marker peak searches will ignore peaks below the threshold value. CALCulate[1|2|3|4|5|6]:TPOWer[:DATA]? Returns the total power of the specified trace. Trace A corresponds to CALCulate1, trace B to CALCulate2, and so on.
book.book Page 67 Monday, January 31, 2000 10:34 AM Remote Operation CALCulate Subsystem Commands CALCulate[1|2|3|4|5|6]:TPOWer:IRANge[STATe] OFF|ON|0|1 CALCulate[1|2|3|4|5|6]:TPOWer:IRANge[STATe]? Turns the total power calculation range for all traces ON or OFF. Setting IRANge:STATe to ON will set the corresponding TPOWer:STATe to ON.
book.book Page 68 Monday, January 31, 2000 10:34 AM Remote Operation CALibration Subsystem Commands CALibration Subsystem Commands CALibration:ALIGn Performs an automatic alignment of the instrument at the wavelength of the largest signal found in full span. This aligns the monochrometer output with the photodetector for improved amplitude accuracy. Sending this command with a marker on screen will generate a “Settings conflict” error. See page 3-3 for additional information on this function.
book.book Page 69 Monday, January 31, 2000 10:34 AM Remote Operation CALibration Subsystem Commands CALibration:POWer Performs a power calibration. The calibration is aborted if the power measured on the input signal is more than 3 dB higher or 10 dB lower than the value specified in the CALibration:POWer:VALue command. CALibration:POWer:DATE? Returns the date of the most recent power calibration.
book.book Page 70 Monday, January 31, 2000 10:34 AM Remote Operation CALibration Subsystem Commands CALibration:WAVelength Performs a wavelength calibration. If the wavelength measured on the input signal differs more than ±2.5 nm from the value specified in the CALibration:WAVelength:VALue command, the calibration is aborted. CALibration:WAVelength:DATE? Returns the date of the most recent wavelength calibration.
book.book Page 71 Monday, January 31, 2000 10:34 AM Remote Operation DISPlay Subsystem Commands DISPlay Subsystem Commands DISPlay[:WINDow[1]]:ANNotation[:ALL] ON|OFF|0|1 DISPlay[:WINDow[1]]:ANNotation[:ALL]? Turns the screen annotation ON or OFF . Affects only the X-axis and Y-axis labeling and labeling within the graticule. DISPlay[:WINDow[1]]:TEXT:CLEar Erases all text on the display resulting from previous use of the DISPlay[:WINDow[1]]:TEXT:DATA command.
book.book Page 72 Monday, January 31, 2000 10:34 AM Remote Operation DISPlay Subsystem Commands Sensitivity dB/div Video Bandwidth Auto Range Enable Trans-Z Lock Repetitive Sweep Sweep Time Auto Chop Mode Gated Sweep Enable Sweep Trigger Mode Trace Length Wavelength Limit Reference Level Linear Display Mode Resolution Bandwidth Res-BW to Span Ratio Peak Search on EOS Line Markers Trace Integration Limit Search Limit Trace Integration Trace Mean Calculation Set according to automeasure setup panel.
book.
book.book Page 74 Monday, January 31, 2000 10:34 AM Remote Operation DISPlay Subsystem Commands DISPlay[:WINDow[1]]:TRACe:X[:SCALe]:AUTO:SPAN [M|NM|UM] DISPlay[:WINDow[1]]:TRACe:X[:SCALe]:AUTO:SPAN? Specifies the final span after a DISPlay:WINDow:TRACe:ALL:SCALe:AUTO command.
book.book Page 75 Monday, January 31, 2000 10:34 AM Remote Operation DISPlay Subsystem Commands DISPlay[:WINDow[1]]:TRACe:Y[1|2][:SCALe]:PDIVision [DB] DISPlay[:WINDow[1]]:TRACe:Y[1|2][:SCALe]:PDIVision? Specifies the dB per division of the vertical scale. Y1 refers to the left (power) scale, and Y2 refers to the right (ratio) scale. The maximum value for the vertical scale is 20 dB per division for the power scale or the ratio scale. The minimum value is 0.01 dB per division.
book.book Page 76 Monday, January 31, 2000 10:34 AM Remote Operation FORMat Subsystem Commands FORMat Subsystem Commands FORMat[:DATA] REAL[,64]|ASCii FORMat[:DATA]? Specifies the trace data format used during data transfer via HP-IB. This command affects data transfers for the TRACe[:DATA] subsystem. The ASCII format is a comma-separated list of numbers. The REAL format is a definitelength block of 64-bit floating-point binary numbers. The definite-length block is defined by IEEE 488.
book.book Page 77 Monday, January 31, 2000 10:34 AM Remote Operation HCOPy Subsystem Commands HCOPy Subsystem Commands HCOPy:DATA? Returns the currently defined printer output as an indefinite length block. After removing the #0 prefix and newline suffix, this block can be saved by the controller and sent directly to a suitable printer. HCOPy:DESTination “SYSTem:COMMunicate:INTernal”|”SYStem:COMMunicate:CENtronics” HCOPy:DESTination? Selects the I/O port for hard copy output.
book.book Page 78 Monday, January 31, 2000 10:34 AM Remote Operation INITiate Subsystem Commands INITiate Subsystem Commands INITiate:CONTinuous OFF|ON|0|1 INITiate:CONTinuous? Specifies repeat or single sweep. INITiate[:IMMediate] Initiates a new measurement sweep.
book.book Page 79 Monday, January 31, 2000 10:34 AM Remote Operation MEMory Subsystem Commands MEMory Subsystem Commands MEMory:STATe[:EXTended]? Returns extended state information as an indefinite length block.
book.book Page 80 Monday, January 31, 2000 10:34 AM Remote Operation MMEMory Subsystem Commands MMEMory Subsystem Commands MMEMory:CATalog? [INTernal|FLOPpy] Lists all files in the current directory.
book.book Page 81 Monday, January 31, 2000 10:34 AM Remote Operation SENSe Subsystem Commands SENSe Subsystem Commands [SENSe:]BANDwidth|BWIDth[:RESolution] [M|NM|UM|A] [SENSe:]BANDwidth|BWIDth[:RESolution]? Sets the resolution bandwidth. Available resolution bandwidths are 10 nm, 5 nm, 2 nm, 1 nm, and either 0.07 nm or 0.06 nm. The narrowest resolution bandwidth in 0.7 nm on the HP 86140A and 86143A. It is 0.06 nm on the HP 86142A and 86145A.
book.book Page 82 Monday, January 31, 2000 10:34 AM Remote Operation SENSe Subsystem Commands [SENSe:]BANDwidth|BWIDth:VIDeo [HZ|KHZ|MHZ|GHZ] [SENSe:]BANDwidth|BWIDth:VIDeo? Specifies the bandwidth of the post-detection video filter. The minimum value for the video bandwidth is 0.1 Hz. The maximum value is the lesser between 3 kHz and the bandwidth of the currently selected transimpedance amplifier. Available choices move in a 1, 3, 10 sequence.
book.book Page 83 Monday, January 31, 2000 10:34 AM Remote Operation SENSe Subsystem Commands [SENSe:]CHOP[:STATe] OFF|ON|0|1 [SENSe:]CHOP[:STATe]? Turns the spectrum analyzer chop mode ON or OFF . Chop mode increases dynamic range for long sweep times by subtracting ambient light. [SENSe:]CORRection:RVELocity:MEDium AIR|VACuum [SENse:]CORRection:RVELocity:MEDium? Determines whether the wavelengths used and displayed are referenced to air or vacuum.
book.book Page 84 Monday, January 31, 2000 10:34 AM Remote Operation SENSe Subsystem Commands The maximum value for Sensitivity is +300 dBm. The minimum value is the value that causes the sweep time to become 1000 seconds, and is an attribute of each individual optical spectrum analyzer. The minimum value will always be less than the values for sensitivity shown in the Specifications section of the User’s Guide.
book.book Page 85 Monday, January 31, 2000 10:34 AM Remote Operation SENSe Subsystem Commands and Span Center = Stop – -------------- """" 2 With Wavelength Limit Off, the minimum value for the Center Wavelength is nominally 350.1 nm. The maximum value is 1999.9 nm. These limits are valid for wavelengths referenced in air or vacuum. With Wavelength Limit On, the minimum value for the Center Wavelength is nominally 600.1 nm. The maximum value is 1699.9 nm.
book.book Page 86 Monday, January 31, 2000 10:34 AM Remote Operation SENSe Subsystem Commands [SENSe:][WAVelength:]SPAN [M|NM|UM|A] [SENSe:][WAVelength:]SPAN? Specifies the wavelength span. The start and stop wavelength and, if necessary, the center wavelength are adjusted so that: Span = 2 ( Center – Start ) and Span = 2( Stop – Center ) The minimum value for Wavelength Span is 0.2 nm. With Wavelength Limit Off, the maximum value for Wavelength Span is 1650 nm.
book.book Page 87 Monday, January 31, 2000 10:34 AM Remote Operation SENSe Subsystem Commands [SENSe:][WAVelength:]SRANge[:STATe] OFF|ON|0|1 [SENSe:][WAVelength:]SRANge[:STATe]? Turns the wavelength sweep range ON or OFF. When the sweep range is ON, the instrument will only sweep between the upper and lower sweep range limits.
book.book Page 88 Monday, January 31, 2000 10:34 AM Remote Operation SENSe Subsystem Commands [SENSe:][WAVelength:]STARt [M|NM|UM|A|HZ|KHZ|MHZ|GHZ] [SENSe:][WAVelength:]STARt? Specifies the start wavelength. The center wavelength and span are adjusted so that: Span Start = Center – -------------2 and Span Stop = Center + -------------2 If the instrument is in zero span, this command sets the center wavelength to the value specified.
book.book Page 89 Monday, January 31, 2000 10:34 AM Remote Operation SENSe Subsystem Commands [SENSe:][WAVelength:]STOP [M|NM|UM|A|HZ|KHZ|MHZ|GHZ] [SENSe:][WAVelength:]STOP? Specifies the stop wavelength. The center wavelength and span are adjusted so that: Span Start = Center – -------------2 and Span Stop = Center + -------------2 If the instrument is in zero span, this command sets the center wavelength to the value specified.
book.book Page 90 Monday, January 31, 2000 10:34 AM Remote Operation SOURce[n] Subsystem Commands SOURce[n] Subsystem Commands SOURce[n]:PULSe:DCYCle SOURce[n]:PULSe:DCYCle? Sets the duty cycle of the sync output. This is equivalent to the TRIGger[:SEQuence]:OUTPut:PULSe:DCYCle command. SOURce[n]:PULSe:WIDTh [S|MS|US|NS] SOURce[n]:PULSe:WIDTh? Sets the pulse width of the sync output. This is equivalent to the TRIGger[:SEQuence]:OUTPut:PULSe:WIDTh command.
book.book Page 91 Monday, January 31, 2000 10:34 AM Remote Operation STATus Subsystem Commands STATus Subsystem Commands STATus:OPERation:CONDition? Queries the contents of the operation condition register. STATus:OPERation:ENABle STATus:OPERation:ENABle? Sets or queries the contents of the operation enable register. STATus:OPERation[:EVENt]? Queries the contents of the operation event register. This query reads the contents of the register and then clears it.
book.book Page 92 Monday, January 31, 2000 10:34 AM Remote Operation STATus Subsystem Commands STATus:QUEStionable:ENABle STATus:QUEStionable:ENABle? Sets or queries the contents of the questionable enable register. STATus:QUEStionable:EVENt? Queries the contents of the questionable event register and then clears it.
book.book Page 93 Monday, January 31, 2000 10:34 AM Remote Operation SYSTem Subsystem Commands SYSTem Subsystem Commands SYSTem:DATE? Queries the date of the real-time clock of the optical spectrum analyzer. SYSTem:ERRor[:NEXT]? Queries the earliest entry in the error queue, thus deleting it. The *CLS command clears the error queue. SYSTem:HELP:HEADers? Returns a list of all commands and queries implemented by the instrument.
book.book Page 94 Monday, January 31, 2000 10:34 AM Remote Operation SYSTem Subsystem Commands SYSTem:TZONe:NAME? Returns the time zone used by the real-time clock of the spectrum analyzer.
book.book Page 95 Monday, January 31, 2000 10:34 AM Remote Operation TRACe Subsystem Commands TRACe Subsystem Commands TRACe[:DATA]:X:STARt? TRA|TRB|TRC|TRD|TRE|TRF Returns the start value for the X-axis data for the trace. The X-axis data will be evenly spaced points from STARt to STOP. The number of points is determined by the TRACe:POINts setting.
book.book Page 96 Monday, January 31, 2000 10:34 AM Remote Operation TRACe Subsystem Commands TRACe[:DATA]:X:[WAVelength]SSTop TRA|TRB|TRC|TRD|TRE|TRF, [M|UM|NM|A|HZ|KHZ|MHZ|GHZ],[M|UM|NM|A|HZ|KHZ|MHZ|GHZ] Sets the start and stop values for the X-axis data for the trace. The first corresponds to the start, and the second corresponds to the stop. If the stop value is a shorter wavelength than the start value, a “Data out of range” error will be generated.
book.book Page 97 Monday, January 31, 2000 10:34 AM Remote Operation TRACe Subsystem Commands TRACe[:DATA][:Y][:POWer] TRA|TRB|TRC|TRD|TRE|TRF, |{,} Sets the Y-axis data points for the trace. The number of Y-axis data points is determined by the TRACe:POINts setting. If a single numeric value is given, all of the Y-axis data points will be set to that value. If more than one value is sent, the trace length will be set to the number of values sent.
book.book Page 98 Monday, January 31, 2000 10:34 AM Remote Operation TRACe Subsystem Commands TRACe:FEED:CONTrol TRA|TRB|TRC|TRD|TRE|TRF,ALWays|NEVer TRACe:FEED:CONTrol? TRA|TRB|TRC|TRD|TRE|TRF Controls how often the specified trace accepts new data. Setting the TRACe:FEED:CONTrol command to ALWays will allow the trace to always accept new data whenever data is available from the FEED. This is equivalent to turning on the trace update from the front panel.
book.book Page 99 Monday, January 31, 2000 10:34 AM Remote Operation TRIGger Subsystem Commands TRIGger Subsystem Commands TRIGger[:SEQuence]:DELay [S|MS|US|NS] TRIGger[:SEQuence]:DELay? Specifies the trigger delay used to start a measurement. TRIGger[:SEQuence]:OUTPut OFF|ON|AUTO TRIGger[:SEQuence]:OUTPut? Controls the ADC trigger output. When OFF the signal will be a TTL LOW. When ON the signal will be a TTL HIGH.
book.book Page 100 Monday, January 31, 2000 10:34 AM Remote Operation TRIGger Subsystem Commands TRIGger[:SEQuence]:SOURce IMMediate|EXTernal|INTernal TRIGger[:SEQuence]:SOURce? Specifies the source, or type, of triggering used to start a measurement. Setting the source to anything other than INTernal sets the slope to EITHer.
book.book Page 101 Monday, January 31, 2000 10:34 AM Remote Operation UNIT Subsystem Commands UNIT Subsystem Commands UNIT:POWer DBM|W|AUTO UNIT:POWer? Specifies the amplitude units for the input, output, and display of the active window and sets the UNIT:RATio units to the corresponding setting. UNIT:RATio DB|LINear|AUTO UNIT:RATio? Specifies units for the input and output of values that represent power ratios.
book.book Page 102 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents HP 71450 Series Commands to HP 86140 Series Equivalents The following table provides a list of the HP 71450 series commands and the SCPI equivalent commands for the HP 86140 series analyzers. Table 4-8.
book.book Page 103 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.book Page 104 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.book Page 105 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.book Page 106 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.book Page 107 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.book Page 108 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.book Page 109 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.book Page 110 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.book Page 111 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.book Page 112 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.book Page 113 Monday, January 31, 2000 10:34 AM Remote Operation HP 71450 Series Commands to HP 86140 Series Equivalents Table 4-8.
book.
book.
book.book Page 2 Monday, January 31, 2000 10:34 AM Status Listings Overview Overview This document describes the status listings of the HP 86140 series optical spectrum analyzers. Status conditions for the optical spectrum analyzer are grouped into categories. Error Indicates the instrument is malfunctioning. Measurement accuracy is probably affected. Errors can be caused by either a hardware or a firmware problem. The instrument requires repair at a Hewlett-Packard service center.
book.book Page 3 Monday, January 31, 2000 10:34 AM Status Listings Overview The following table lists the error numbers and their definitions.
book.book Page 4 Monday, January 31, 2000 10:34 AM Status Listings Error Reporting Behavior Error Reporting Behavior Errors are displayed in an on-screen dialog box. To continue operation, the user must acknowledge the error by pressing a button. Status errors are displayed with a descriptive line in the lower-left corner of the graticule. Immediately press the System key below the display, and then press the Help softkey to the right of the display to display more information.
book.book Page 5 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors SCPI-Defined Errors These error messages and descriptions were copied from the SCPI 1997 Volume 2: Command reference. The sentences enclosed in brackets “[ ]” are copied from the error descriptions in the SCPI reference. References are also made to IEEE 488.2 sections for further clarification of events. Not all the available SCPI error messages were utilized in the optical spectrum analyzer.
book.book Page 6 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors Contact: The Institute of Electrical and Electronics Engineers, Inc. 345 East 47th Street New York, New York 10017-2394 Phone: (800) 678-IEEE (US) 8 a.m. – 4:30 p.m. (EST) (908) 981-1393 (International) Fax: (908) 981-9667 Standard SCPI errors (–1 to –999) All positive numbers are instrument-dependent.
book.book Page 7 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors listening formats or whose type is unacceptable to the device. • An unrecognized header was received. Unrecognized headers include incorrect device-specific headers and incorrect or unimplemented IEEE 488.2 common commands. • A Group Execute Trigger (GET) was entered into the input buffer inside of an IEEE 488.2 .
book.book Page 8 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors Table 5-1. Command Errors (2 of 4) Error Number Error Description [description/explanation/examples] –108 desc = “Parameter not allowed” help = ““ [More parameters were received than expected for the header; for example, the *EMC common command only accepts one parameter, so receiving *EMC 0,1 is not allowed.
book.book Page 9 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors Table 5-1. Command Errors (3 of 4) Error Number Error Description [description/explanation/examples] –128 desc = “Numeric data not allowed” help = ““ [A legal numeric data element was received, but the device does not accept one in this position for the header.] –131 desc = “Invalid suffix” help = ““ The suffix does not follow the syntax described in IEEE 488.2, 7.7.3.
book.book Page 10 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors Table 5-1. Command Errors (4 of 4) Error Number Error Description [description/explanation/examples] –161 desc = “Invalid block data” help = ““ [A block data element was expected, but was invalid for some reason (see IEEE 488.2, 7.7.6.2); for example, an END message was received before the length was satisfied.
book.book Page 11 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors Execution errors An in the range [–299 , –200] indicates that an error has been detected by the instrument’s execution control block. The occurrence of any error in this class shall cause the execution error bit (bit 4) in the event status register (IEEE 488.2, section 11.5.1) to be set.
book.book Page 12 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors Table 5-2. Execution Errors (2 of 3) Error Number Error Description [description/explanation/examples] –222 desc = “Data out of range” help = “A numeric value was entered which is outside the legal range of values for the parameter. The name of the parameter is listed at the end of the error message.
book.book Page 13 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors Table 5-2. Execution Errors (3 of 3) Error Number Error Description [description/explanation/examples] –273 desc = “Illegal macro label” help = ““ [Indicates that the macro label defined in the *DMC command was a legal string syntax, but could not be accepted by the device (see IEEE 488.2, 10.7.3 and 10.7.6.
book.book Page 14 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors 488.2, section 11.5.1) to be set. Events that generate device-specific errors shall not generate command errors, execution errors, or query errors; see the other error definitions in this section. Table 5-3. Device-Specific Errors Error Number Error Description [description/explanation/examples] –310 desc = “System error” help = ““ [Indicates that some error, termed “system error” by the device, has occurred.
book.book Page 15 Monday, January 31, 2000 10:34 AM Status Listings SCPI-Defined Errors Table 5-4. Query Errors Error Number Error Description [description/explanation/examples] –400 Query error [This is the generic query error for devices that cannot detect more specific errors. This code indicates only that a Query Error as defined in IEEE 488.2, 11.5.1.1.7 and 6.3 has occurred.] –410 Query INTERRUPTED [Indicates that a condition causing an INTERRUPTED Query error occurred (see IEEE 488.2, 6.3.2.
book.book Page 16 Monday, January 31, 2000 10:34 AM Status Listings OSA Notices OSA Notices System control-related error messages or warnings The OSA system changed a setting and generated a warning that the operation was performed. Table 5-5.
book.book Page 17 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings OSA Warnings Table 5-6. OSA Warnings (1 of 11) Error Number Error Description [description/explanation/examples] 5000 desc = “AutoMeasure cannot find an input signal” help = “The auto-measure procedure cannot find a usable input signal. Make sure you have a signal connected to the optical input. Auto-measure will not work with very small input signals. They must be measured manually.
book.book Page 18 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings Table 5-6. OSA Warnings (2 of 11) Error Number Error Description [description/explanation/examples] 5005 desc = “Cal aborted: amplitude correction too large” help = “An amplitude calibration was requested. The calibration was aborted since the correction needed is more than +3dB or less than –10dB. Make sure you have done an Auto-Align prior to calibration.
book.book Page 19 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings Table 5-6. OSA Warnings (3 of 11) Error Number Error Description [description/explanation/examples] 5012 desc = “The floppy disk is not formatted” help = “An operation was requested which uses the floppy disk. The disk currently in the drive does not appear to be formatted. Use the format utility to format the disk or replace it with a formatted disk.
book.book Page 20 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings Table 5-6. OSA Warnings (4 of 11) Error Number Error Description [description/explanation/examples] 5032–5043 desc = “I/O Error” help “An unexpected error occurred during the I/O operation. Please try the operation again. If the operation involves the floppy disk drive, try a different floppy disk.
book.book Page 21 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings Table 5-6. OSA Warnings (5 of 11) Error Number Error Description [description/explanation/examples] 5049 desc = “Wrong marker X axis units for active trace” help = “The active marker cannot be placed on the active trace because the desired X axis units do not match the X axis units of the active trace. This is normally caused by recalling a wavelength trace in zero span or a zero span trace in a wavelength span.
book.book Page 22 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings Table 5-6. OSA Warnings (6 of 11) Error Number Error Description [description/explanation/examples] 5055 desc = “Firmware Upgrade was not successful” help => “A firmware upgrade operation was requested. The firmware upgrade operation cannot be done at this time due to an internal software problem. “ + <.
book.book Page 23 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings Table 5-6. OSA Warnings (7 of 11) Error Number Error Description [description/explanation/examples] 5062 desc = “ADC Triggered Sweep Too Fast” help = “A sweep was taken with one of the ADC trigger modes enabled. The sweep rate was too fast to allow trace data to be acquired for every wavelength. To resolve this problem increase sweep time or increase the frequency the external trigger input signal.
book.book Page 24 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings Table 5-6. OSA Warnings (8 of 11) Error Number Error Description [description/explanation/examples] 6722 desc = “Math expression input cannot be boolean.” help = “A math expression could not be evaluated because one or more of the input arguments is boolean. This function does not allow boolean input arguments.” 6723 desc = “Trace X axis values do not match.
book.book Page 25 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings Table 5-6. OSA Warnings (9 of 11) Error Number Error Description [description/explanation/examples] 6731 desc = “Illegal combination of trace Y axis units” help = “A math expression could not be evaluated because the combination of Y axis units in the input arguments do not make sense.
book.book Page 26 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings Table 5-6. OSA Warnings (10 of 11) Error Number Error Description [description/explanation/examples] 6739 desc = “Too many peaks or pits” help = “A marker search operation was requested. There are too many valid pits or peaks in the current trace. Reduce the number of valid pits or peaks by increasing the peak or pit excursion setting.
book.book Page 27 Monday, January 31, 2000 10:34 AM Status Listings OSA Warnings Table 5-6. OSA Warnings (11 of 11) Error Number Error Description [description/explanation/examples] 6750 desc = “Pit not found” help = “A pit search was requested. There are no valid pits on the current marker’s trace. Try decreasing the marker pit excursion setting or adjust the sensitivity setting.” 6751 desc = “Trace has no centroid” help = “The trace has no centroid.
book.book Page 28 Monday, January 31, 2000 10:34 AM Status Listings Application-Specific Warnings Application-Specific Warnings Table 5-7. Application-Specific Warnings (1 of 6) Error Number Error Description [description/explanation/examples] 8001 desc = “Incorrect application type is listed in spec file.” help = “The application expects the first non-comment line of the specification file to contain the APPLICATION keyword followed by the application type.
book.book Page 29 Monday, January 31, 2000 10:34 AM Status Listings Application-Specific Warnings Table 5-7. Application-Specific Warnings (2 of 6) Error Number Error Description [description/explanation/examples] 8006 desc = “The specification file cannot be imported.” help = “An error occurred while trying to import the specification file. Refer to the previous warnings for more information on specific errors in the specification file.” 8007 desc = “The specification file has a syntax error.
book.book Page 30 Monday, January 31, 2000 10:34 AM Status Listings Application-Specific Warnings Table 5-7. Application-Specific Warnings (3 of 6) Error Number Error Description [description/explanation/examples] 8014 desc = “Print statement ignored: no path is specified” help = “The PRINT statement needs to be after a PATH statement to indicate which PATH data is to be printed. The PRINT statement at the line number specified came before any PATH statement.
book.book Page 31 Monday, January 31, 2000 10:34 AM Status Listings Application-Specific Warnings Table 5-7. Application-Specific Warnings (4 of 6) Error Number Error Description [description/explanation/examples] 8021 desc = “The spec file could not be found.” help = “The application tried to load a specification file which could not be found in the internal memory. This may have occurred if the last file used by the application was deleted. The filename is specified within the parentheses.
book.book Page 32 Monday, January 31, 2000 10:34 AM Status Listings Application-Specific Warnings Table 5-7. Application-Specific Warnings (5 of 6) Error Number Error Description [description/explanation/examples] 8028 desc = “A minimum non-zero span is required.” help = “The application cannot run in a zero span setting. The start and stop wavelength must be separated by a minimum span.
book.book Page 33 Monday, January 31, 2000 10:34 AM Status Listings Application-Specific Warnings Table 5-7. Application-Specific Warnings (6 of 6) Error Number Error Description [description/explanation/examples] 8035 desc = “The search limits are outside the SETUP range.” help = “The search limits for statements like PEAK or CENTER_OF_MASS must be within the start and stop values of the SETUP statement. The line number indicates which statement has values out of range.
book.book Page 34 Monday, January 31, 2000 10:34 AM Status Listings OSA Status Errors OSA Status Errors Table 5-8. OSA Status Errors Error Number Error Description [description/explanation/examples] 10000 desc = “Sweep Uncalibrated” help = “The current setting of sweep time may be too fast. This could result in an invalid measurement. In certain cases it is possible to sweep faster than the coupled sweep time without compromising measurement accuracy.
book.book Page 35 Monday, January 31, 2000 10:34 AM Status Listings OSA Errors OSA Errors Table 5-9. OSA Errors (1 of 2) Error Number Error Description [description/explanation/examples] 20001 desc = “Error detected in ADC sub-system” help = “An error has been detected in the Analog-to-Digital converter subsystem. Please record the hexadecimal number listed with the error and cycle power. If the error persists, contact the nearest Hewlett-Packard Instrument support center for assistance. In the U.S.
book.book Page 36 Monday, January 31, 2000 10:34 AM Status Listings OSA Errors Table 5-9. OSA Errors (2 of 2) Error Number Error Description [description/explanation/examples] 20005 desc = “Trans-Impedance calibration data is invalid” help = “Factory calibration data for the trans-impedance amplifier is invalid. Please record the hexadecimal number listed with the error and cycle power. If the error persists, contact the nearest Hewlett-Packard Instrument support center for assistance. In the U.S.
book.book Page 37 Monday, January 31, 2000 10:34 AM Status Listings Firmware Errors Firmware Errors Table 5-10. Firmware Errors Error Number Error Description [description/explanation/examples] 30000 desc = “Internal Communications Error” help = “An internal software error has occurred involving communications between different software processes. Please record this error including the extra text and cycle power.
book.
book.
book.book Page 2 Monday, January 31, 2000 10:34 AM Reference Options and Accessories Options and Accessories Option/HP Part Number Description HP 86140A HP 86142A HP 86143A HP 86145A Table 6-1.
book.book Page 3 Monday, January 31, 2000 10:34 AM Reference Options and Accessories Option/HP Part Number Description HP 86143A HP 86145A 9211-5604 Transit Case ✔ ✔ Option 042 Soft Carrying Case ✔ ✔ HP N1031A BenchLink Softwarec (included) ✔ ✔ ✔ HP 86142A HP 86140A Table 6-1. Options and Accessories (2 of 2) ✔ a. Contact your local field representative for availability. b. Always use HP paper. Use of other thermal paper may degrade image quality and damage the printer head. c.
book.book Page 4 Monday, January 31, 2000 10:34 AM Reference AC Line-Power Cords AC Line-Power Cords Plug Type Cable Part No.
book.book Page 5 Monday, January 31, 2000 10:34 AM Reference AC Line-Power Cords Plug Type Cable Part No. Plug Description Length (in/cm) Color Country 100V 8120-4753 Straight MITI 90/230 Dark Gray Japan 8120-4754 90° 90/230 * Part number shown for plug is the industry identifier for the plug only. Number shown for cable is the HP part number for the complete cable including the plug.
book.
book.book Page 7 Monday, January 31, 2000 10:34 AM Reference Printer Head Cleaning Procedure Printer Head Cleaning Procedure Lint from normal use of the printer may eventually collect on the printer head and degrade print quality. Use the procedure provided in this section to clean the printer head. WARN IN G This servicing procedure is for use by qualified personnel only. To avoid electrical shock, do not perform this procedure unless you are qualified to do so.
book.book Page 8 Monday, January 31, 2000 10:34 AM Reference Printer Head Cleaning Procedure Figure 6-1. Example of a static-safe workstation Procedure 1 Turn off the HP 86140-series optical spectrum analyzer, and remove the line power cord. 2 Place the instrument at a static-safe work station as described in the introduction to this procedure. 3 Use a coin or screwdriver to open the printer door that is located on the top of the instrument.
book.book Page 9 Monday, January 31, 2000 10:34 AM Reference Printer Head Cleaning Procedure 5 Unscrew the retaining screw that secures the sheet-metal cover that protects the printer head from electrostatic discharge. Slide the sheet-metal cover towards the retaining screw and then lift it straight up to remove. 6 Lift the printer head lever to the vertical position. Then, tilt the lever towards the instrument’s rear panel to rotate the printer head up.
book.book Page 10 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements Cleaning Connections for Accurate Measurements Advances in measurement capabilities make connectors and connection techniques more important than ever. Damage to the connectors on calibration and verification devices, test ports, cables, and other devices can increase downtime and expense.
book.book Page 11 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements Choosing the Right Connector A critical, but often overlooked, factor in making a good lightwave measurement is the selection and care of the fiber-optic connector. The differences in connector types are mainly in the mechanical assembly that holds the ferrule in position against another identical ferrule. Connectors also vary in the polish, curve, and concentricity of the core within the cladding.
book.book Page 12 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements Universal adapters to Diamond HMS_10. The HMS-10 encases the fiber within a soft nickel silver (CuNiZn) center which is surrounded by a tough tungsten carbide casing, as shown in Figure 6-2. Staking Groove (Fixing oper.) Secondary Staking (Active C entering) Nic kel Silver (Cu/Ni/Zn) (Soft center) Tungsten Carbide (Hard Case) 125 um Fiber (Centered to about 0.2 microns) Figure 6-2.
book.book Page 13 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements Inspecting Connectors Because fiber-optic connectors are susceptible to damage that is not immediately obvious to the naked eye, bad measurements can be made without the user even being aware of a connector problem. Although microscopic examination and return loss measurements are the best way to ensure good connections, they are not always practical.
book.book Page 14 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements Figure 6-4. A dirty endface from poor cleaning Repeated connections made without removing loose particles, or using improper cleaning tools can lead to physical damage of the glass fiber endface, as shown in Figure 6-4. When severe, the damage on one connector end can be transferred to another good connector that comes in contact with it. Figure 6-5.
book.book Page 15 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements Guidelines Use the following guidelines to achieve the best possible performance when making measurements on a fiber-optic system: • Keep connectors covered when not in use. • Use dry connections whenever possible. • Never use metal or sharp objects to clean the connector and never scrape the connector. • Avoid matching gel and oils.
book.book Page 16 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements • Use fusion splices on the more permanent critical nodes. Choose the best connector possible. Replace connecting cables regularly. Frequently measure the return loss of the connector to check for degradation, and clean every connector, every time. All connectors should be treated like the high-quality lens of a good camera.
book.book Page 17 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements Visual inspection of fiber ends Visual inspection of fiber ends can be helpful. Contamination or imperfections on the cable end can be detected, as well as cracks or chips in the fiber itself. Use a microscope with 100× to 200× magnification to inspect the entire end face for contamination, raised metal, or dents in the metal as well as any other imperfections.
book.book Page 18 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements Table 6-3. Cleaning Accessories Item HP Part Number Isopropyl alcohol 8500-5344 Cotton swabs 8520-0023 Small foam swabs 9300-1223 Compressed dust remover (non-residue) 8500-5262 Table 6-4.
book.book Page 19 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements 5 Immediately dry the fiber end with a clean, dry, lint-free cotton swab or lens paper. 6 Blow across the connector end face from a distance of 6 to 8 inches using filtered, dry, compressed air. Aim the compressed air at a shallow angle to the fiber end face. Nitrogen gas or compressed dust remover can also be used. CAUTION Do not shake, tip, or invert compressed air canisters.
book.book Page 20 Monday, January 31, 2000 10:34 AM Reference Cleaning Connections for Accurate Measurements 3 Dry the inside of the adapter with a clean, dry foam swab. 4 Blow through the adapter using filtered, dry, compressed air. CAUTION Do not shake, tip, or invert compressed air canisters. This releases particles from the can into the air. Refer to the instructions on the compressed air canister.
book.book Page 21 Monday, January 31, 2000 10:34 AM Reference Returning the Instrument for Service Returning the Instrument for Service The instructions in this section show you how to properly package the instrument for return to a Hewlett-Packard service office. For a list of offices, refer to “Hewlett-Packard Sales and Service Offices” on page 6-24.
book.book Page 22 Monday, January 31, 2000 10:34 AM Reference Returning the Instrument for Service information should be returned with the instrument. • Type of service required. • Date instrument was returned for repair. • Description of the problem: • Whether problem is constant or intermittent. • Whether instrument is temperature-sensitive. • Whether instrument is vibration-sensitive. • Instrument settings required to reproduce the problem. • Performance data. • Company name and return address.
book.book Page 23 Monday, January 31, 2000 10:34 AM Reference Returning the Instrument for Service Sealed Air Corporation (Commerce, California 90001). Air Cap looks like a plastic sheet filled with air bubbles. Use the pink (antistatic) Air Cap™ to reduce static electricity. Wrapping the instrument several times in this material will protect the instrument and prevent it from moving in the carton. 4 Seal the carton with strong nylon adhesive tape. 5 Mark the carton “FRAGILE, HANDLE WITH CARE”.
book.book Page 24 Monday, January 31, 2000 10:34 AM Reference Hewlett-Packard Sales and Service Offices Hewlett-Packard Sales and Service Offices Before returning an instrument for service, call the HP Instrument Support Center at (800) 403-0801. Hewlett-Packard Sales and Service Offices (1 of 2) U.S. FIELD OPERATIONS Headquarters Hewlett-Packard Company 19320 Pruneridge Avenue Cupertino, CA 95014 U.S.A.
book.book Page 25 Monday, January 31, 2000 10:34 AM Reference Hewlett-Packard Sales and Service Offices Hewlett-Packard Sales and Service Offices (2 of 2) INTERCON FIELD OPERATIONS Headquarters Hewlett-Packard Company 3495 Deer Creek Rd. Palo Alto, California 94304-1316 (415) 857-5027 Australia Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130 (61 3) 895-2895 Canada Hewlett-Packard Ltd.
book.
book.
book.book Page 2 Monday, January 31, 2000 10:34 AM Specifications and Regulatory Information Specifications and Regulatory Information Specifications and Regulatory Information This chapter contains specifications and characteristics for HP 86140 series optical spectrum analyzers. The specifications in this chapter apply to all functions autocoupled over the temperature range 0°C to +55°C and relative humidity < 95% (unless otherwise noted).
book.book Page 3 Monday, January 31, 2000 10:34 AM Specifications and Regulatory Information Definition of Terms Definition of Terms Characteristics The distinction between specifications and characteristics is described as foland specifications lows: Specifications describe warranted performance. Characteristics provide useful, but nonwarranted information about the functions and performance of the instrument. Characteristics are printed in italics.
book.book Page 4 Monday, January 31, 2000 10:34 AM Specifications and Regulatory Information Definition of Terms Amplitude Sensitivity is defined as the signal level that is equal to six times the RMS value of the noise. Displayed sensitivity values are nominal. Slightly lower values may have to be entered to achieve specified sensitivity. Scale Fidelity refers to the potential errors in amplitude readout at amplitudes other than at the calibration point. This specification is sometimes called linearity.
book.book Page 5 Monday, January 31, 2000 10:34 AM Specifications and Regulatory Information Specifications Specifications HP 86140A and HP 86143A HP 86142A and HP 86145A WAVELENGTH Range 600 nm to 1700 nm 600 nm to 1700 nm Span Range (continuously variable) 0.2 nm to full range and zero span 0.2 nm to full range and zero span ±0.025 nm (1510–1570), ±0.035 nm (1570–1640) ±0.05 nm ±0.025 nm (1510–1570), ±0.035 nm (1570–1640) ±0.05 nm ±0.2 nm ±0.2 nm ±0.5 nm ±0.
book.book Page 6 Monday, January 31, 2000 10:34 AM Specifications and Regulatory Information Specifications HP 86140A and HP 86143A HP 86142A and HP 86145A AMPLITUDE Sensitivity d 750 to 900 nm (no averaging required) e –60 dBm –75 dBm –60 dBm –75 dBm 900 to 1250 nm (no averaging required) e –75 dBm –75 dBm 1250 to 1610 nm (no averaging required) e –90 dBm –90 dBm b –80 dBm –80 dBm +15 dBm +15 dBm +12 dBm +15 dBm +15 dBm +12 dBm +30 dBm +23 dBm +30 dBm +23 dBm ±0.5 dB ±0.5 dB ±0.
book.book Page 7 Monday, January 31, 2000 10:34 AM Specifications and Regulatory Information Specifications HP 86140A and HP 86143A HP 86142A and HP 86145A DYNAMIC RANGE In 0.1 nm resolution a,k 1250 to 1610 nm (chop mode on) at ±0.5 nm, ±1 nm, ±5 nm 1550 nm at ±0.8 nm (±100 GHz at 1550 nm) 1550 nm at ±0.5 nm (±62.5 GHz at 1550 nm) 1550 nm at ±0.4 nm (±50 GHz at 1550 nm) 1550 nm at ±0.
book.book Page 8 Monday, January 31, 2000 10:34 AM Specifications and Regulatory Information Specifications HP 86140A and HP 86143A HP 86142A and HP 86145A COMPUTER INTERFACING Remote Control Compatibility Interfaces IEEE-488.1, IEEE-488.2 (100%) HP-IB, Parallel Printer Port, External VGA Monitor, Keyboard (PS/2) and Mouse 3.5 inch 1.
book.book Page 9 Monday, January 31, 2000 10:34 AM Specifications and Regulatory Information Specifications General Specifications HP 86140A and HP 86142A Benchtop HP 86143A and HP 86145A Portable 222 mm H × 425 mm W × 427 mm D (8.8 in × 16.8 in × 16.8 in) 16.5 kg (36 lb) Dimensions Weight 163 mm H × 325 mm W × 427 mm D (6.4 in × 12.8 in × 16.8 in) 14.
book.book Page 10 Monday, January 31, 2000 10:34 AM Specifications and Regulatory Information Regulatory Information Regulatory Information • Laser Classification: This product contains an FDA Laser Class I (IEC Laser Class 1) laser. • This product complies with 21 CFR 1040.10 and 1040.11. Notice for Germany: Noise Declaration Acoustic Noise Emission Geraeuschemission LpA < 70 dB LpA < 70 dB Operator position am Arbeitsplatz Normal position normaler Betrieb per ISO 7779 nach DIN 45635 t.
book.
book.
book.book Page 1 Monday, January 31, 2000 10:34 AM Index A absolute accuracy, 7-3 ac line-power cords, 6-4 accessories, 6-2 connecting, 1-5 static-safe, 6-7 accuracy absolute/differential, 7-3 amplitude, 3-2 active area, moving, 3-47 function area assist, 3-44 marker, automatic measurement, 3-54 marker, setting, 3-23 trace, setting, 3-23, 3-57 Active Marker.... softkey, 3-23, 3-26–3-27 Active Trace....
book.book Page 2 Monday, January 31, 2000 10:34 AM Index Center WL softkey, 3-62 characteristics, 7-2–7-9 Choose File to Recall or Delete softkey, 3-37 Choose Files to Save softkey, 3-36 classification, laser, 7-10 cleaning connections, 6-10 printer head, 6-7 Close Panel....
book.book Page 3 Monday, January 31, 2000 10:34 AM Index error messages Bandwidth not found, 3-24 settings conflict, 4-9 Exchange Menu....
book.book Page 4 Monday, January 31, 2000 10:34 AM Index line markers integrating, 3-30 searching between, 3-29 sweep between, 3-29 wavelength, 3-28 Line Markers Off softkey, 3-29 line power cords, 6-4 requirements, 1-7 Local key, 3-5 LOCAL softkey, 4-4 long form commands, 4-7 lowercase letters in commands, 4-7 M marker, 3-54 line, searching between, 3-29 line, sweep between, 3-29 search threshold, setting, 3-33 settings, default, 3-30 types, 2-22 Marker BW....
book.book Page 5 Monday, January 31, 2000 10:34 AM Index OPTICAL INPUT connector, 2-3 optimize sensitivity, 3-54 options, 6-2 Options menu, 2-10 OSA Backup Utility screen, 3-39 display, 2-8 Restore Utility screen, 3-39 State information panel, 3-47 OSA Extended State.... softkey, 3-53 OSA State....
book.book Page 6 Monday, January 31, 2000 10:34 AM Index remote interface automatic alignment, 3-3 automatic measurement, 3-4 Remote Setup.... softkey, 3-55 repeat sweep, 3-17 Repeat Sweep softkey, 3-17 repeatability, tuning, 7-3 reproducibility, 7-3 Res BW softkey, 3-15 resolution bandwidth, 3-15 FWHM, 7-3 restore internal memory, 3-39 Restore Utility screen, 3-39 return loss, connectors, 6-16 returning data, 4-10 Revision.... softkey, 3-42 S safety information, iii Save Menu....
book.
book.book Page 8 Monday, January 31, 2000 10:34 AM Index start/stop, 3-62 step size, 3-64 units, 3-63 Wavelength Cal Setup.... softkey, 3-45 Wavelength functions center wavelength, 3-62 centering the peak wavelength, 3-63 default settings, 3-63 span, 3-62 start wavelength, 3-62 stop wavelength, 3-62 Wavelength Limit softkey, 3-51 Wavelength Line Mkr softkey, 3-28 Wavelength menu, 2-19, 3-61 Wavelength Setup....