specifications 8753ET/ES Network Analyzers HP 8753ET and 8753ES Network Analyzers 30 kHz to 3 or 6 GHz This document describes the performance and features of the following products: HP 8753ES S-parameter vector network analyzer HP 8753ES Option 011 vector network analyzer without test set HP 8753ET Transmission/reflection vector network analyzer For more information about these analyzers, please see the following documents: HP 8753ET and 8753ES Network Analyzers Brochure HP 8753ET and 8753ES Network A
Table of contents Definitions and test conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 System performance summaries . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Specifications and characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Measurement throughput summary . . . . . . . . . . . . . . . . . . . . . . . . .21 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Definitions and test conditions Specifications describe the instrument's warranted performance after a half-hour warm-up and over the temperature range of 25 ˚C ± 5 ˚C, unless otherwise stated. Specifications for frequencies above 3 GHz do not apply to instruments with Option 075 (75-ohm impedance). Supplemental characteristics are typical but non-warranted performance parameters. These are denoted as “typical,” “nominal,” or “approximate.
System performance summaries Measurement port characteristics HP 8753ES (50-ohm systems) 7-mm test ports Frequency range Corrected 30 kHz-300 kHz2 300 kHz-1.3 GHz 1.3 GHz-3 GHz Directivity 55 dB 55 dB 51 dB Source match 55 dB 51 dB 49 dB Load match 55 dB 55 dB 51 dB Reflection ±(0.001 dB ±(0.001 dB ±(0.005 dB tracking +0.02 dB/˚C) +0.01 dB/˚C) +0.02 dB/˚C) Transmission ±(0.008 dB ±(0.006 dB ±(0.009 dB tracking +0.02 dB/˚C) +0.01 dB/˚C) +0.
System performance summaries Measurement port characteristics HP 8753ES (50-ohm systems) Type-N test ports Frequency range Corrected 30 kHz-300 kHz2 300 kHz-1.3 GHz 1.3 GHz-3 GHz Directivity 50 dB 50 dB 47 dB Source match 49 dB 42 dB 36 dB Load match 50 dB 50 dB 47 dB Reflection ±(0.005 dB ±(0.009 dB ±(0.019 dB tracking +0.02 dB/˚C) +0.01 dB/˚C) +0.02 dB/˚C) Transmission ±(0.014 dB ±(0.013 dB ±(0.026 dB tracking +0.02 dB/˚C) +0.01 dB/˚C) +0.
System performance summaries HP 8753ES (50-ohm systems) 3.5-mm test ports Configuration Network analyzer HP 8753ES Standard and Option 006 HP 85033D HP 11857D Calibration kit Test-port cables System dynamic range Measurement port characteristics Frequency range Corrected 30 kHz-300 kHz2 300 kHz-1.3 GHz 1.3 GHz-3 GHz 3 GHz-6 GHz Directivity 49 dB 46 dB 44 dB 38 dB Source match 49 dB 44 dB 41 dB 37 dB Load match 49 dB 46 dB 44 dB 38 dB Reflection ±(0.010 dB ±(0.005 dB ±(0.007 dB ±(0.009 dB tracking +0.
System performance summaries Measurement port characteristics HP 8753ES (75-ohm systems) Type-N test ports Frequency range Corrected 30 kHz - 300 kHz2 300 kHz - 1.3 GHz 1.3 GHz - 3 GHz Directivity 48 dB 48 dB 43 dB Source match 47 dB 41 dB 35 dB Load match 48 dB 48 dB 43 dB Reflection ±(0.004 dB ±(0.010 dB ±(0.019 dB tracking +0.02 dB/˚C) +0.01 dB/˚C) +0.02 dB/˚C) Transmission ±0.018 dB ±(0.016 dB ±(0.033 dB tracking +0.02 dB/˚C) +0.01 dB/˚C) +0.02 dB/˚C) Uncorrected2 30 kHz - 300 kHz 300 kHz - 1.
System performance summaries HP 8753ES (75-ohm systems) Type-F test ports Configuration Network analyzer Calibration kit Test-port cables HP 8753E Option 075 HP 85039B HP 11857B System dynamic range 68 dB1 90 dB1 96 dB 106 dB 30 kHz to 50 kHz 50 kHz to 300 kHz 300 kHz to 16 MHz 16 MHz to 3 GHz Measurement port characteristics Data is shown for a Type-F female reflection port and a Type-F male transmission port.
System performance summaries HP 8753ET (50-ohm systems) Type-N test ports Configuration Network analyzer Calibration kit Test port cable HP 8753ET Standard, Option 006 or Option 004 HP 85032B HP part number 8120-4781 System dynamic range 300 kHz to 16 MHz 16 MHz to 3 GHz 3 GHz to 6 GHz 100 dB 110 dB 105 dB Measurement port characteristics Frequency range (continued) Corrected 300 kHz-1.3 GHz 1.3 GHz-3.0 GHz 3.0 GHz-6.
HP 8753ES specifications Test port input characteristics 30 kHz to 3 GHz Option 006 30 kHz to 6 GHz Frequency range Test-port output characteristics7 Frequency characteristics Range 30 kHz to 3 GHz Option 006 Resolution Stability9 Accuracy Option 1D5 Power range2 Option 075/014 30 kHz to 6 GHz 1 Hz ±7.5 ppm (0˚ to 55 ˚C) ±3 ppm/year ±10 ppm at 25 ˚C ± 5 ˚C ±0.05 ppm (0˚ to 55 ˚C), ±0.
Test port input characteristics (continued) External source mode1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements (typical) Power level 0 to -25 dBm Spectral purity Maximum <-30 dBc spurious input Residual FM <20 kHz Typical settling 500 ms (automatic) time Reference level Range Resolution Stability2 30 kHz to 3 GHz 3 GHz to 6 GHz ±500 dB 0.001 dB 0.02 dB/ ˚C 0.
HP 8753ES Option 011 specifications Test port input characteristics Option 011 300 kHz to 3 GHz Option 006 30 kHz to 6 GHz Frequency range Test port output characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability typically ±7.5 ppm 0˚ to 55 ˚C typically ±3 ppm/year Option 1D5 typically ±0.05 ppm 0˚ to 55 ˚C typically ±0.5 ppm/year Accuracy ±10 ppm at 25 ˚C ± 5 ˚C Power range -5 to +20 dBm Option 006 -5 to +18 dBm Resolution 0.
HP 8753ES Option 011 specifications Test port input characteristics (continued) Frequency offset mode3 Frequency range Option 006 300 kHz to 3 GHz 300 kHz to 6 GHz R-channel input requirements Power level 300 kHz to 3 GHz 3 GHz to 6 GHz 0 to -35 dBm 0 to -30 dBm LO spectral purity Maximum spurious <-25 dBc input Residual FM <20 kHz LO frequency accuracy ±1 MHz at nominal frequency (CW time sweep only) 300 kHz to 6 GHz External source mode4 Frequency range R-channel input requirements1 Power level 0
HP 8753ET specifications Test-port output characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz Resolution 1 Hz Stability4 ±7.5 ppm (0˚ to 55 ˚C) Option 1D5 Accuracy Power range: Resolution Level accuracy1 Level linearity1 Standard ±3 ppm/year ±0.05 ppm (0˚ to 55 ˚C) ±0.5 ppm/year ±10 ppm at 25 ˚C ±5 ˚C -20 to +5 dBm -85 to +10 dBm (with Option 004) 0.01 dB ±1.0 dB -20 to -15 dBm -15 to 0 dBm 0 to +5 dBm ±0.5 dB ±0.2 dB ±0.
HP 8753ET specifications Test-port input characteristics (continued) External source mode1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements2 Power level 0 to -25 dBm Spectral purity Maximum spurious <-30 dBc input <20 kHz Residual FM Typical settling time 500 ms (automatic) Frequency readout accuracy2 Input frequency accuracy Manual: Accuracy Reference level Range ±500 dB (300 kHz to 3 GHz) 50 ms (manual) 0.1% (automatic) requirement2 -0.
HP 8753ET/ES supplemental characteristics Measurement Accuracy The following graph shows group-delay accuracy at 1.3 GHz with type-N full two-port calibration and 10-Hz IF bandwidth. Insertion loss is assumed to be < 2 dB and electrical length to be ten meters. Number of display channels Four display channels available. Number of measurement channels Two primary and two auxiliary measurement channels available. Measurement parameters HP 8753ET: Reflection, transmission, A, B, R, A/R, B/R, A/B.
HP 8753ET/ES supplemental characteristics (continued) Sweep trigger Set to either continuous, single, group sweep, or external trigger. Set external trigger to take a complete sweep or to measure individual points in a frequency, power or list sweep. Power Set source power from -20 to +5 dBm for the HP 8753ET1 or from -85 to +10 dBm for the HP 8753ES2. Power slope can be set in dBm/GHz. Power meter calibration Select continuous leveling or use a correction table to modify source power.
HP 8753ET/ES supplemental characteristics (continued) Trace functions Display data Display current measurement data, memory data, or current measurement with measurement and memory data simultaneously. Trace math Vector division or subtraction of current linear measurement values and memory data.
HP 8753ET/ES supplemental characteristics (continued) • Low-pass impulse Frequency offset on/off • Bandpass impulse Sets the RF source to be swept at a fixed offset frequency above the receiver as required in a swept RF/IF, fixed LO, mixer test. The bandpass impulse stimulates a pulsed RF signal (with an impulse envelope) and is used to measure the time-domain response of band-limited devices. Service menu Windows Select the desired service test, service diagnostic, service or verification mode.
General characteristics Front panel connectors HP 8753ES test ports (without Option 011) Connector type 7 mm, precision Impedance 50 ohms (nominal) Test sequence output (TEST SEQ) By default, this connector outputs a TTL end-of-sweep signal. It can also be programmed by the user in a test sequence to output a user-defined TTL signal. BNC (f) connector.
Environmental characteristics Measurement throughput summary General conditions RFI and EMI susceptibility: defined by VDE 0730, CISPR Publication 11, and FCC Class B Standards. The following table shows typical measurement times in milliseconds. ESD (electrostatic discharge): must be eliminated by use of static-safe work procedures and an anti-static bench mat. The flexible rubber keypad protects key contacts from dust, but the environment should be as dust-free as possible for optimal reliability.
Test set specifications for Option 011 HP 85046A/B S-parameter test sets The HP 85046A/B S-parameter test sets provide the capability to measure reflection and transmission characteristics (including S-parameters) of two-port devices in either direction with a single connection. The test sets are controlled from the HP 8753ES Option 011 and include a programmable step attenuator. The frequency range of the HP 85046A 50-ohm test set is 300 kHz to 3 GHz. The HP 85046A has precision 7-mm connectors.
HP 85047A S-parameter test set The HP 85047A S-parameter test set provides the capability to simultaneously measure the reflection and transmission characteristics of two-port devices in either direction with a single connection. This test set includes a frequency doubler that can be switched in by an HP 8753B/C Option 006 to measure 3 MHz to 6 GHz in a single sweep or switched out to measure 300 kHz to 3 GHz in a single sweep.
Accessories HP 85033D 3.5-mm calibration kit Calibration kits Contains a set of precision 3.5-mm standards to calibrate the HP 8753ES and 50-ohm test sets for the measurement of devices with precision 3.5-mm and SMA connectors. Precision phase-matched 7-mm to 3.5-mm adapters are included for accurate measurements of non-insertable devices. Option 001 deletes the 7-mm to type-N adapters.
HP 85038 family of calibration kits HP 11906B 7-16 to 50-Ohm type-N adapter kit Description The HP 85038A calibration kit contains open circuits, short circuits, and terminations with both male and female 7-16 connectors. More economical versions of this kit are available: the HP 85038M contains only male devices, and the HP 85038F provides only female devices (see following table).
Verification kits HP 11850C/D three-way power splitters Measuring known devices other than the standards used in calibration is an easy way to verify the proper operation of an HP 8753ES measurement system. HP offers verification kits which include devices, with data, for verifying the error-corrected measurements of an HP 8753ES and 50-ohm test sets.
HP 11852B 50-ohm to 75-ohm minimum loss pad (continued) Connectors Dimensions Weight 50-ohm type-N (f) to 75-ohm type-N (m) standard Option 004, 50-ohm type-N (m) to 75-ohm type-N (f) 14-mm D x 70-mm L (0.56 in x 2.75 in) Net 0.1 kg (0.316 lb) RF limiter Externally attaches to one or both ports of the analyzer. Provides protection against potential high-power transients from external devices.
HP 8347A RF amplifier The HP 8347A RF amplifier delivers increased power across a 300 kHz to 3 GHz frequency range. The HP 8347A provides leveled output power without using an external coupler and detector, since these parts are built-in. The external ALC can be directly connected to the external AM input on the HP 8753ET/ES. This capability is especially useful for achieving high dynamic range measurements at faster sweep rates.