User's Manual

I
-
I
-
Spectrum Analyzer Mode Key Descriptions
Spectrum Analyzer Mode Functions
SRC PWR
ON OFF
SRC PWR
STP SIZE
START
F=Q
STATE
+
CARD
Option 011
only.
activates (ON) or deactivates (OFF) the output power of
the tracking generator (SRC). The power level can then be
adjusted using the data keys, step keys, or knob. Pressing
any digit, 0 through 9, on the numeric keypad brings up
the selected terminator menu. The available output power
level varies for different analyzer models. See specifications
and characteristics in your calibration guide for specific
information for your analyzer.
Front-Panel Key Access
IJiZ5Z~
Option 011 only.
allows the user to set the step size of the source-power
level, source-power offset, and power-sweep range functions.
The step size may be values from -32.7
dB
to 32.7
dB.
The
default setting is one vertical scale division.
Front-Panel Key Access
(AUX]
sets the frequency at the left side of the graticule. The left
and right sides of the graticule correspond to the start and
stop frequencies. When these frequencies are activated,
their values are displayed below the graticule in place of
center frequency and span.
Front-Panel Key Access (FREQUENCY)
saves the current analyzer state on the memory card. To
save the current state, press STATE
-+
CARD , use the
numeric keypad to enter a number, and press ENTER If
you want the file name of the stored data to contain a
prefix, press Change Prefix to enter a prefix before
storing the data. If the state data was stored using a prefix,
the hle name for the state data consists of s(prefix)-(register
number). If no prefix was specified, the hle name is
s-(register number). If windows are being used, only the
state of the active window will be saved.
Front-Panel Key Access
m
6-137