User's Manual

564 LOG AMPL
565 LOG AMPL
566 LOG AMPL
567 LOG AMPL
568 LOG AMPL
569 LOG AMPL
570 LOG AMPL
571 AMPL
572 AMPL
IM
573 LOG AMPL
574 LOG AMPL
575 LOG AMPL
576 LOG AMPL
577 LOG AMPL
578 LOG AMPL
579 LOG AMPL
580 LOG AMPL
581 AMPL
Unable to adjust amplitude of log scale. No compression in
CalLogPower .
Unable to adjust amplitude of log scale. Gain compression error in
CalLogPower.
Unable to adjust amplitude of log scale. Unable to set LOG CAL LVL
in CalLogPower
Unable to adjust amplitude of log scale. No TOS LX in
CalLogExpand. Possible problem in Log offset/Log Expand stage.
Unable to adjust amplitude of log scale. No LVL LX in CalLog Power.
Possible problem in Log offset/Log Expand stage.
Unable to adjust amplitude of log scale. Low MDACX value in
CalLogExpand. Possible problem in Log offset/Log Expand stage.
Unable to adjust amplitude of log scale. High MDACX value in
CalLogExpand. Possible problem in Log offset/Log Expand stage.
Unable to adjust step gain amplifiers. No
TOS-D
in
CalDetectors.
Unable to adjust amplitude of 1 MHz Resolution Bandwidth. No
TOS-W in
CalWidebandLog.
Unable to adjust amplitude of log scale. Video offset error
+O.
Check
video offset circuitry.
Unable to adjust amplitude of log scale. Video offset error
t.1.
Check
video offset circuitry.
Unable to adjust amplitude of log scale. Video offset error
+2.
Check
video offset circuitry.
Unable to adjust amplitude of log scale. Video offset error
+3.
Check
video offset circuitry.
Unable to adjust amplitude of log scale. Video offset error
+4.
Check
video offset circuitry.
Lim Calibration error from DC Logger Cal
Attenuator CAL level error from DC Logger Cal.
FID CAL level error from DC Logger Cal
Unable to adjust 100 kHz and
510
kHz Resolution Bandwidths.
ADC/CALOSC
handshake CAL
in Sweep Xtal problem. Refer to
Error 582.
General Troubleshooting 6-29