bx="0.150" by="0.060" wx="0.3127" wy="0.329" wY="80"
external="FALSE"/>
</calibration_info>
</calibration>
</studiocal>
Tento druhý príklad dáva monitoru pokyn na kalibráciu dvoch predvolieb, prvej s primárnymi zložkami BT.
709, bielym bodom D65 a gamou BT.1886. Druhá sa bude kalibrovať s primárnymi zložkami DCI-P3, bielym
bodom P3 a gamou 2,6. Obe kalibrácie ukladá na miesto pamäte od výrobcu.
<?xml version="1.0" encoding="ISO-8859-1"?>
<!--BT.709/BT.1886 calibration-->
<studiocal>
<calibration entries="2">
<calibration_info>
<preset num="3" name="BT.709" store="factory"/>
<gamma value="bt1886"/>
<target_primaries rx="0.640" ry="0.330" gx="0.300" gy="0.600"
bx="0.150" by="0.060" wx="0.3127" wy="0.329" wY="80"
external="FALSE"/>
</calibration_info>
<calibration_info>
<preset num="5" name="DCI P3" store="factory"/>
<gamma value="2.6"/>
<target_primaries rx="0.680" ry="0.320" gx="0.265" gy="0.690"
bx="0.150" by="0.060" wx="0.314" wy="0.351" wY="48"
use_uv="false" external="false"/>
</calibration_info>
</calibration>
</studiocal>
Generovanie prolu ICC
V niektorých pracovných postupoch sa vyžaduje prol ICC kalibrovanej predvoľby. Tento prol môžete
generovať pomocou tagu save_icc. Tento tag dáva monitoru pokyn na výpočet prolu ICC pre kalibrovanú
predvoľbu a jeho zápis na jednotku USB ash obsahujúcu súbor StudioCal XML. Výpočet prolu ICC vyžaduje,
aby sa vykonali ďalšie meraniach po kalibrácii, takže kalibrácie pomocou spektrorádiometra môžu trvať
výrazne dlhšie.
Tento tag obsahuje jeden atribút a je štruktúrovaný nasledovne:
<save_icc value="TRUE|FALSE"/>
Základné kalibračné tagy 81