Instructions

HT9014 - HT9015
EN - 9
4.3.2. DC Voltage measurement
CAUTION
Maximum input for DC Voltage measurements is 1000V. Do not take any
voltage measurement exceeding this limit in order not to risk electrical
shock or damaging the tester.
Fig. 4: Taking DC voltage measurements
1. Rotate the switch on position
2. Insert the red test lead plug into V
CAPHz%Temp jack and the black test lead
plug into COM jack (Fig. 4)
3. Connect the two long ends of test leads to the desired circuit, then reading will be
displayed
4. When O.L." symbol is displayed, the result is over the maximum value that the
instrument is able to measure
5. For HOLD, MAX/MIN and REL features please refer to § 4.2
CAUTION
Due to the high input impedance is possible that the meter performs the
zero of the display taking longer
The possible blinking values shown at display with open input terminals
are not a problem of the meter and these values will not be added to
the measured value