Instructions

HT9014 - HT9015
EN - 19
AC TRMS Current (instrument HT9014)
Range Resolution Accuracy (*) Bandwidh Overload protection
60.00A 0.01A
±(2.8%rdg+12dgt)
50 ÷ 60Hz
600AACrms
600.0A 0.1A
±(2.8%rdg+8dgt)
60.00A 0.01A
±(4.5% rdg+10dgt)
61 ÷ 400Hz
600.0A 0.1A
±(5.0%rdg+10dgt)
PEAK function: response time <10ms ; (*) Referred to cable inside to the center of clamp jaws
Position sensitivity: ±2.0%rdg
AC TRMS Current (instrument HT9015)
Range Resolution Accuracy (*) Bandwidh Overload protection
60.00A 0.01A
±(2.2%rdg+12dgt)
50 ÷ 60Hz
600AACrms
600.0A 0.1A
±(2.2%rdg+8dgt)
60.00A 0.01A
±(3.5% rdg+12dgt)
61 ÷ 400Hz
600.0A 0.1A
±(3.5%rdg+8dgt)
PEAK function: response time <10ms ; (*) Referred to cable inside to the center of clamp jaws
Position sensitivity: ±2.0%rdg
Capacitance (Autorange)
Range Resolution Accuracy Overload protection
40.00nF 0.01nF
±(3.5%rdg+40dgt)
600VDC/ACrms
400.0nF 0.1nF
±(2.5%rdg+5dgt)
4.000µF 0.001µF
40.00µF 0.01µF
400.0µF 0.1µF
4000µF 1µF ±(5.0%rdg+5dgt)
Diode test
Range Test current Open voltage
0.9mA typical 2.8VDC
Frequency with test leads (Autorange)
Range Resolution Accuracy Sensitivity Overload protection
99.99Hz 0.01Hz
±(1.0%rdg+5dgt) 15Vrms
600VDC/ACrms
999.9Hz 0.1Hz
9.999kHz 0.001kHz
60.00kHz 0.01kHz
Frequency with jaws (Autorange)
Range Resolution Accuracy Sensitivity Overload protection
99.99Hz 0.01Hz
±(1.0%rdg+5dgt)
10A (60A)
50A(600A)
600AACrms
999.9Hz 0.1Hz
9.999kHz 0.001kHz
Duty Cycle (Autorange)
Range Resolution Sensitivity Accuracy
0.5% ÷ 99.0%
0.1%
15Vrms ±(1.2%rdg+2dgt)
Temperature with type K probe (Autorange)
Range Resolution Accuracy (*) Overload protection
-20.0 ÷ 400.0°C
0.1°C
±(2.0%rdg +3°C)
600VDC/ACrms
400 ÷ 760°C
1°C
±(2.0%rdg +5°C)
-4.0 ÷ 752.0°F
0.1°F
±(2.0%rdg +6°F)
752 ÷ 1400°F
1°F
±(2.0%rdg +9°F)
(*) Accuracy of type K probe not considered